Propaq CS Directions for Use
57
Alarms &
Aler
ts
Troubleshooting Equipment Alert Messages
ECG Messages
If a lead fault occurs, the ECG equipment alert typically shows which lead failed. If multiple leads fail,
the monitor displays MULTIPLE.
ECG LEAD CHANGED
. The Propaq CS monitor has automatically changed an ECG lead due to a
lead wire or electrode problem.
LEAD FAIL: REPLACE ELECTRODES
. The cable may not be properly connected to the
electrodes or the electrodes may have failed. Check for proper connection; replace electrodes if
needed.
RESP Messages
LEAD FAIL.
One or more electrodes are making very poor or no contact. Check for proper
connection; replace electrodes if needed.
INAPPROPRIATE ECG CABLE.
ECG cable appears not to contain 1 k
Ω
current limiting resistors.
These resistors are required for RESP operation and to protect the monitor from damage during
defibrillation. Replace cable with proper type.
NOISY SIGNAL, CHECK ELECTRODES.
Electrodes are making poor contact and may be dried
out. Replace electrodes.
IBP Messages
TRANSDUCER NOT DETECTED.
The transducer connection is broken.
TRANSDUCER SHORT CIRCUIT.
This message appears when the Propaq CS monitor senses a
short in the transducer. The transducer should be replaced.
INCOMPATIBLE TRANSDUCER.
Check the compatible transducers listed in the Protocol
Systems Products and Accessories booklet to confirm you are using a compatible transducer.
NIBP Messages
If an error number (ERR# x) is listed in an NIBP trend printout or display, it indicates that the
corresponding NIBP equipment alert occurred.
AIR LEAK, CHECK HOSE
(ERR# 1). The Propaq CS monitor could not properly inflate cuff.
Check the hose and cuff for obvious leaks, such as the O-rings in the hose connections.
CUFF NOT DETECTED
(ERR# 2). During cuff inflation the detected pressure did not sufficiently
rise. Check that the cuff connection is tight and take the measurement again.
KINKED HOSE, CHECK HOSE
(ERR# 3). The Propaq CS monitor could not properly inflate cuff.
Check for a kinked hose between the monitor and the patient.
Содержание Propaq 242
Страница 6: ...6 Protocol Systems Inc...
Страница 18: ...18 Protocol Systems Inc...
Страница 26: ...26 Protocol Systems Inc...
Страница 50: ...50 Protocol Systems Inc...
Страница 52: ...52 Protocol Systems Inc...
Страница 64: ...64 Protocol Systems Inc...
Страница 66: ...66 Protocol Systems Inc...
Страница 72: ...72 Protocol Systems Inc...
Страница 74: ...74 Protocol Systems Inc...
Страница 78: ...78 Protocol Systems Inc...
Страница 84: ...84 Protocol Systems Inc...
Страница 86: ...86 Protocol Systems Inc...
Страница 94: ...94 Protocol Systems Inc...
Страница 120: ...120 Protocol Systems Inc...