Protimeter Grainmaster   

Page 3

5.

Ensure that the grinder-compressor unit is clean and is operating smoothly. If not,

disassemble and clean as detailed in section 5.1.

6.

Prior to placing the grinder-compressor unit on the instrument, ensure that the plunger

is fully retracted within its housing. The plunger is retracted fully by sliding the switch

on the underside of the rotating handle forward into the compress position (picture)

and rotating handle anti-clockwise until the clutch clicks a few times. The sliding

switch can only be pushed or pulled into position when the black pips on the top of the

rotating handle are aligned with the yellow X.

7.

Pull the switch on the underside of the

rotating handle to the grind position and

place the grinder-compressor unit over

the grain cup. Lock the unit in place by

twisting it clockwise against the three

lugs.

8.

Always use the correct sample size of 10

ml. Using the 10 ml spoon provided, pour

a sample into the hopper on the side of

the grinder-compressor unit.

9.

Hold the 

Grainmaster i

 against a flat and horizontal surface and rotate the handle

clockwise to grind the sample and push it through into the grain cup. If necessary,

check to see the entire sample has passed through the grinder blade by turning the

handle anti-clockwise a quarter turn and looking into the hopper.

10.

Align the pips and the yellow X of the rotating handle and push the sliding switch

forward into the compress position. Turn the handle clockwise to screw the plunger

onto the ground sample that is now in the grain cup. Correct compression is reached

when the clutch clicks a few times.

11.

Press and hold 

 to display the moisture content of the sample. When the reading has

stabilized, release 

 to freeze the

%H

2

O

 value for approximately 7 seconds. Note

and record this value as required.

12.

If necessary, display the temperature of the sample by pressing the 

 or 

 

only

 after

releasing 

. Note and record this value as required.

13.

Having noted the moisture (and/or temperature) reading, turn the handle anti-

clockwise until the clutch clicks (to fully retract the plunger!). Remove the grain cup

and the tested sample and clean the grain cell with the brush prior to commencing

another test.

Содержание Grainmaster i

Страница 1: ...vanced Sensors PROTIMETER Grainmaster i Protimeter Grain Moisture Meter Instruction Manual 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipme...

Страница 2: ......

Страница 3: ...INS3000 Rev B June 2014 PROTIMETER Amphenol Advanced Sensors Grainmaster i Protimeter Grain Moisture Meter Instruction Manual...

Страница 4: ......

Страница 5: ...he Pre Programmed Calibrations 4 3 Optional External Moisture and Temperature Probes 4 3 1 Using the Optional Moisture and Temperature Probe No GRN3005 5 3 2 Using the Optional Temperature Probe Part...

Страница 6: ...cting the operational modes and taking measurements There is a socket on the left hand side of the instrument for optional probes see section 3 for details The instrument is supplied with the followin...

Страница 7: ...s are displayed 2 Measuring Moisture and Temperature of Ground and Compressed Samples To take moisture and temperature measurements of 10 ml samples of crops complete the following steps 1 Ensure that...

Страница 8: ...or unit 9 Hold the Grainmaster i against a flat and horizontal surface and rotate the handle clockwise to grind the sample and push it through into the grain cup If necessary check to see the entire s...

Страница 9: ...g the button to display the H2O value press to increase the calibration or to decrease the calibration in increments of 0 1 An flashes in the bottom right corner of the display to indicate that a cali...

Страница 10: ...e instrument as described above 3 Press and release to switch ON confirm that the instrument senses external moisture probe flagged by display showing PROBE if not press once to do so 4 As with the in...

Страница 11: ...ll toggle back to the internal grain cell Note There is no user adjustment to the temperature reading 3 3 Using the Optional Bale Probes Part Nos GRN6138 and GRN6138 S Note An optional Bale Probe can...

Страница 12: ...ion The buttons have the following effects Exit setup mode without any further changes Save any changes and advance to next option or Modify the value for the selected option Example To change tempera...

Страница 13: ...regular intervals especially when testing wet or oily crops that tend to clog the thread of the plunger spindle Follow the maintenance procedure as detailed in the following steps 1 Disengage the two...

Страница 14: ...rned to Protimeter for servicing 6 Reactivate ATC as detailed in section 4 0 before continuing to use the instrument 7 Information Codes The instrument displays a range of codes that represent various...

Страница 15: ...he method of use of the instrument and its accessories and the interpretation of the readings are beyond the control of the manufacturers they cannot accept responsibility for any loss consequential o...

Страница 16: ...nol Advanced Sensors Copyright 2014 Amphenol Thermometrics Inc 967 Windfall Road St Marys PA 15857 3333 USA 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us a...

Отзывы: