Protimeter Grainmaster   

Page 7

4 Setup  Mode

The setup mode is entered from the switched – 

OFF

 state by pressing 

 while switching 

ON

 using 

. This action displays the firmware version of the instrument (example 1.00)

 

until all buttons are released. The product part number (example Grn3000)

 

then scrolls 

across the display from right to left followed by the firmware date in 

yy-mm-dd

 format 

(example 

00-06-16

)

 

and then the first set up code, 

0=0

.

The user then has the option of restoring the factory settings (

°C, ATC

 

ON

, all user crop 

adjustments cleared) or simply changing the default setting for the temperature display (

°C 

or °F

) or of enabling/ disabling the automatic temperature correction.

The buttons have the following effects:

Exit setup mode without any further changes.

 Save any changes and advance to next option

 or 

 Modify the value for the selected option.

Example:

 To change temperature the display from 

°C 

to

 °F

 and switch 

OFF ATC

:

1.

Press and hold 

, briefly press 

, release both buttons.

2.

Wait until 

0=0

 is displayed, press 

 to display 

1=0

°C

3.

Press 

 to change display to 

1=1

°F

 (Temperatures now in 

°F

)

4.

Press 

 to save this change and display 

2=0

Table 3: Setup Mode Options

Display

Description

0=0

No action

0=1

Resets all user settings to the default settings (

°C, ATC

 

ON

).

Also resets all user crop adjustments to zero.

1=0

Selects 

°C

 for temperature display

 

(accompanied by °

C

 on display)

1=1

Selects 

°F

 for temperature display (accompanied by 

°F

 on display)

2=0

Activates automatic temperature correction (

!

 in corner cleared)

2=1

Deactivates temperature correction (

!

 displays)

Содержание Grainmaster i

Страница 1: ...vanced Sensors PROTIMETER Grainmaster i Protimeter Grain Moisture Meter Instruction Manual 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipme...

Страница 2: ......

Страница 3: ...INS3000 Rev B June 2014 PROTIMETER Amphenol Advanced Sensors Grainmaster i Protimeter Grain Moisture Meter Instruction Manual...

Страница 4: ......

Страница 5: ...he Pre Programmed Calibrations 4 3 Optional External Moisture and Temperature Probes 4 3 1 Using the Optional Moisture and Temperature Probe No GRN3005 5 3 2 Using the Optional Temperature Probe Part...

Страница 6: ...cting the operational modes and taking measurements There is a socket on the left hand side of the instrument for optional probes see section 3 for details The instrument is supplied with the followin...

Страница 7: ...s are displayed 2 Measuring Moisture and Temperature of Ground and Compressed Samples To take moisture and temperature measurements of 10 ml samples of crops complete the following steps 1 Ensure that...

Страница 8: ...or unit 9 Hold the Grainmaster i against a flat and horizontal surface and rotate the handle clockwise to grind the sample and push it through into the grain cup If necessary check to see the entire s...

Страница 9: ...g the button to display the H2O value press to increase the calibration or to decrease the calibration in increments of 0 1 An flashes in the bottom right corner of the display to indicate that a cali...

Страница 10: ...e instrument as described above 3 Press and release to switch ON confirm that the instrument senses external moisture probe flagged by display showing PROBE if not press once to do so 4 As with the in...

Страница 11: ...ll toggle back to the internal grain cell Note There is no user adjustment to the temperature reading 3 3 Using the Optional Bale Probes Part Nos GRN6138 and GRN6138 S Note An optional Bale Probe can...

Страница 12: ...ion The buttons have the following effects Exit setup mode without any further changes Save any changes and advance to next option or Modify the value for the selected option Example To change tempera...

Страница 13: ...regular intervals especially when testing wet or oily crops that tend to clog the thread of the plunger spindle Follow the maintenance procedure as detailed in the following steps 1 Disengage the two...

Страница 14: ...rned to Protimeter for servicing 6 Reactivate ATC as detailed in section 4 0 before continuing to use the instrument 7 Information Codes The instrument displays a range of codes that represent various...

Страница 15: ...he method of use of the instrument and its accessories and the interpretation of the readings are beyond the control of the manufacturers they cannot accept responsibility for any loss consequential o...

Страница 16: ...nol Advanced Sensors Copyright 2014 Amphenol Thermometrics Inc 967 Windfall Road St Marys PA 15857 3333 USA 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us a...

Отзывы: