Page 8

Protimeter Grainmaster

5.

Press 

 to change display to 

2=1

!

 (

ATC

 is now deactivated.)

6.

Press 

to save this change and display 

0=0

7.

Press 

 to exit.

Note:

It is recommended that the 

ATC

 is only be deactivated when using the ‘

Quickcheck’

 

to verify the calibration of the instrument (see section 6.0). Whenever 

ATC

 is 

OFF

 the 

°C

 

or 

°F

 icon (as appropriate) is flashed as a warning while moisture measurements 

are displayed.

5 Care and Maintenance

When not in use, store the 

Grainmaster i

 in a stable, dust-free environment and out of 

direct sunlight. Remove the battery from the instrument if it is to be stored for periods of 

more than four weeks, or when the low battery power symbol appears on the display. 

Check the condition of accessories used with the instrument on a regular basis and replace 

them if they become worn or damaged.

5.1 Grinder-Compressor

The Grinder-Compressor unit should be cleaned and lubricated at regular intervals, 

especially when testing wet or oily crops that tend to clog the thread of the plunger spindle. 

Follow the maintenance procedure as detailed in the following steps:

1.

Disengage the two halves of the Grinder-Compressor unit by twisting the yellow

locking ring counterclockwise and pulling apart.

2.

Open the blade retaining wings and lift out the plunger and blade assembly.

3.

Separate the blade from the plunger assembly and remove the yellow feeder ring from

inside the top half of the Grinder-Compressor unit.

4.

Clean the blade, feeder ring and both halves of the Grinder-Compressor unit with the

brush.

5.

Clean the thread of the plunger unit and ensure it spins freely. Lubricate with a light

oil.

6.

Replace the feeder ring in the top half of the Grinder-Compressor unit.

7.

Put the blade on the plunger assembly and reposition this unit in the bottom half of the

Grinder-Compressor unit. Clamp in place by closing the retaining wings.

8.

Slide the two halves of the Grinder-Compressor together, ensuring the location lugs

and plunger spindle are correctly aligned.

9.

Replace the yellow locking ring.

Содержание Grainmaster i

Страница 1: ...vanced Sensors PROTIMETER Grainmaster i Protimeter Grain Moisture Meter Instruction Manual 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipme...

Страница 2: ......

Страница 3: ...INS3000 Rev B June 2014 PROTIMETER Amphenol Advanced Sensors Grainmaster i Protimeter Grain Moisture Meter Instruction Manual...

Страница 4: ......

Страница 5: ...he Pre Programmed Calibrations 4 3 Optional External Moisture and Temperature Probes 4 3 1 Using the Optional Moisture and Temperature Probe No GRN3005 5 3 2 Using the Optional Temperature Probe Part...

Страница 6: ...cting the operational modes and taking measurements There is a socket on the left hand side of the instrument for optional probes see section 3 for details The instrument is supplied with the followin...

Страница 7: ...s are displayed 2 Measuring Moisture and Temperature of Ground and Compressed Samples To take moisture and temperature measurements of 10 ml samples of crops complete the following steps 1 Ensure that...

Страница 8: ...or unit 9 Hold the Grainmaster i against a flat and horizontal surface and rotate the handle clockwise to grind the sample and push it through into the grain cup If necessary check to see the entire s...

Страница 9: ...g the button to display the H2O value press to increase the calibration or to decrease the calibration in increments of 0 1 An flashes in the bottom right corner of the display to indicate that a cali...

Страница 10: ...e instrument as described above 3 Press and release to switch ON confirm that the instrument senses external moisture probe flagged by display showing PROBE if not press once to do so 4 As with the in...

Страница 11: ...ll toggle back to the internal grain cell Note There is no user adjustment to the temperature reading 3 3 Using the Optional Bale Probes Part Nos GRN6138 and GRN6138 S Note An optional Bale Probe can...

Страница 12: ...ion The buttons have the following effects Exit setup mode without any further changes Save any changes and advance to next option or Modify the value for the selected option Example To change tempera...

Страница 13: ...regular intervals especially when testing wet or oily crops that tend to clog the thread of the plunger spindle Follow the maintenance procedure as detailed in the following steps 1 Disengage the two...

Страница 14: ...rned to Protimeter for servicing 6 Reactivate ATC as detailed in section 4 0 before continuing to use the instrument 7 Information Codes The instrument displays a range of codes that represent various...

Страница 15: ...he method of use of the instrument and its accessories and the interpretation of the readings are beyond the control of the manufacturers they cannot accept responsibility for any loss consequential o...

Страница 16: ...nol Advanced Sensors Copyright 2014 Amphenol Thermometrics Inc 967 Windfall Road St Marys PA 15857 3333 USA 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us a...

Отзывы: