V4.0_2021-05-26
PROGNOST C Instructions for use
5073-0-0002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
12 von 30
2.1.4
Pinching and Collision Hazards
CAUTION!
Ensure that while using any product that can be lowered, raised or moved in
different directions, neither yourself (operator), the patient or any third party
finds themselves in a hazardous position (area of movement). Remove all
objects (e.g. chairs, pushcarts) from known collision areas.
Be aware that careless or improper adjustment of the radiographic system
(movement of column, detector Bucky, Vertical Bucky wall stand and table
top) can lead to damage of the X-Ray components, unusable X-Ray images
and injury to the patient. Failure to pay attention can lead to damage of the
radiographic system as well as external objects.
2.1.5
Explosion protection
These PROGNOST C is not designated for use within areas with explosive hazards.
2.1.6
Interaction with external devices
Unwanted interaction with external devices is not known.
2.1.7
Electromagnetic Environment and the influence of devices
CAUTION!
The use of other accessories, other converters and other cables than those
specified by PROTEC or provided in the documentation of the component
manufacturer can result in increased electromagnetic interference or reduced
electromagnetic immunity of the device and lead to defective operation.
CAUTION!
The use of the PROGNOST C immediately next to other devices or with other
devices in a stacked form should be avoided, as this could result in defective
operation. If use in the manner described above is nevertheless necessary, the
PROGNOST C and the other devices should be observed to ensure that they
are working properly.
NOTE
The properties of this device, determined by emissions, allow its use in industrial
areas and in hospitals (CISPR 11, class A). When used in residential areas (for which
class B is usually required by CISPR 11), this device may not provide adequate
protection for radio services. The user may need to take remedial measures such as
relocating or realigning the device.
The PROGNOST C is intended for use in an environment in professional health care facilities (e.g., clinics,
surgery centres, physiology practices ...)