PROGNOST XP, XPE, XPE-Akku
Instructions for use
5038-0-0002
21 of 51
3
3x
Overload, excessive patient load
and thus excessive current
consumption of the drives
Reduce the patient load and
allow the device to cool
down at a standstill. If the
cause has been eliminated,
the status messages can be
deleted by briefly pressing
the enable key
4
4x
Unintentional movement, e.g.,
downwards due to
impermissibly high patient load
Reduce the patient load.
Notify PROTEC authorized
service
only
XPE-Akku
5x
DC-DC converter defect
Height adjustment not
possible. Take the device out
of order and notify PROTEC
authorized service
only
XPE-Akku
6x
Battery empty
Height adjustment not
possible. Replace or charge
battery
5
7x
The control has detected that
the upper or lower limit switch
is out of function
Take the device out of order
and notify PROTEC
authorized service
8x
Height difference of the lifting
columns, there is 1 cm
difference in the position
(height) of the lifting columns
•
Perform zero balance
•
Notify a PROTEC
authorized service if the
error message reoccurs
after the zero balance
9X
Blocking of the table top brake
Take the device out of order
and notify PROTEC
authorized service
10X
Duty cycle of the lifting
columns exceeded
Allow the device to cool
down at standstill until
signal LED changes from red
flashing rhythm to bright
green again
NOTE
State message 2,4 and 5
:
If these state messages can be deleted with the release button and then occur
again, the PROTEC authorized service must be notified and the table must be taken
out of order.