3.3.3
Emergency Stop Switch and Status Indication
The control unit and the drive for the height adjustment are switched off by actuating the emergency
stop switch.
The emergency stop switch is unlocked by turning it to the right.
Next to the emergency stop there is the three-color signal lamp, which is used to indicate readiness
and status messages.
WARNING!
Even if the emergency stop switch has been actuated and the signal lamp is
not illuminated, the mains voltage may still be present at the device plug. The
table is only safely disconnected from the power supply when the power plug
is pulled out of the supply socket.
The following is an overview of the signal lamps status indication:
Indicator light color
Status
Product concerned
slight green
Energy saving mode
o
Operational readiness is activated by
pressing the release key
PROGNOST XPE-Akku
green
Device is ready for operation
PROGNOST XPE &
PROGNOST XPE-Akku
orange
Battery almost empty
o
Normal driving function still possible
PROGNOST XPE-Akku
red
Battery empty
o
Only showing driving function
possible
o
No restart in energy saving mode
possible
o
Battery must be replaced
PROGNOST XPE-Akku
Error message
PROGNOST XPE &
PROGNOST XPE-Akku
Status indications for all XP-series models are given by cyclic flashing of the indicator light in the color
red.
State
Flashing-Rhythm red
Description
Measure
1
1x
Over temperature of power
amplifier.
Caused by to frequent up and
down moving or too heavy
patient load
Cool down the device in
standby until red blinking
cycle is finished and
indicator lamp is bright
green again.
Reduce the patient load
2
2x
Drive blocked
Check if there is something
between the table top and
the floor, move the table up
and remove the object.
Notify PROTEC authorized
service.