Protec 7036-0-1710 Скачать руководство пользователя страница 1

 

PROTEC GmbH & Co. KG 

In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 
Telephone: +49 (0) 7062/92 55-0 E-Mail: [email protected]

 

Version: 1.0 

Issued: 2021-05-17 

Subject to alterations

 

 

 
 
 

 
 

PROGNOST XP, XPE & XPE-Akku 

 

Basic diagnostic X-ray system table,  

non-powered or powered 

 

 

Model/ID: 7036-0-17xx

 

Basic UDI-DI: 426050264X009ZT 

Model/ID: 7037-x-17xx

 

Basic UDI-DI: 426050264X011ZE 

Model/ID: 7038-9-17xx 

Basic UDI-DI: 426050264X024ZP 

 

 

Instructions for use 

 

ID No.  5038-0-0002 

 

 

*Figure PROGNOST XPE-Akku 

Содержание 7036-0-1710

Страница 1: ...sued 2021 05 17 Subject to alterations PROGNOST XP XPE XPE Akku Basic diagnostic X ray system table non powered or powered Model ID 7036 0 17xx Basic UDI DI 426050264X009ZT Model ID 7037 x 17xx Basic...

Страница 2: ...ited without the express written consent of PROTEC s law department Knowledge of violations of these regulations must be reported immediately to PROTEC GmbH Co KG 2021 PROTEC GmbH Co KG Oberstenfeld C...

Страница 3: ...eral Safety Instructions 13 2 1 1 Requirements for operation 13 2 1 2 Device Operation 13 2 1 2 1 Operating Type 13 2 1 3 Operating Personnel 13 2 1 4 Crushing and Collision Hazard 13 2 1 4 1 PROGNOST...

Страница 4: ...2 3 Specifications for PROGNOST XPE Akku Operation 26 4 2 3 1 Energy Saving Mode 26 4 2 3 2 Battery Box 26 4 2 3 3 Battery Charge Level 28 4 2 3 4 Battery Box Charging Station 28 4 2 3 5 Charging the...

Страница 5: ...e 40 8 Description of Symbols Labels and Abbreviations 41 8 1 Symbols 41 8 2 Type Label 43 8 2 1 Type Label PROGNOST XP 43 8 2 2 Type Label PROGNOST XPE 43 8 2 3 Type Label PROGNOST XPE Akku 43 8 3 La...

Страница 6: ...the configuration of the equipment as of the date of manufacture Revisions to the equipment subsequent to the date of manufacture will be addressed in service updates distributed to the PROTEC Technic...

Страница 7: ...of these accompanying documents is required to carefully read through and carefully consider the instructions warnings and cautions contained therein before starting operation Even if you have already...

Страница 8: ...automatic exposure control unit The floating table top is locked in horizontal and vertical direction with a mechanical brake and can be released by a brake bracket The wheels of the table can be loc...

Страница 9: ...hort 2 pieces which is mounted on the front of the table top to facilitate the displacement of the table or the table top Side protection bar for Z angle to protect the detector Fixing mechanism Shock...

Страница 10: ...the table top prepared to attach several table accessories Suitable for digital Bucky 1 3 2 X ray System Table PROGNOST XPE PROGNOST XPE Akku Floating tabletop Table top color white Motorized height...

Страница 11: ...no intended main effect in or on the human body Therefore considered in isolation no indications or contraindications can be shown for them 1 9 Intended User Group As a component of a diagnostic X ray...

Страница 12: ...ntained therein must be carefully read and adhered to NOTE After the initial installation the commissioning must be recorded with the PROTEC acceptance protocol FB 04 07A4 NOTE The PROGNOST XP product...

Страница 13: ...uous operation of 1 5 minutes 2 1 3 Operating Personnel NOTE Only trained and authorized personnel are allowed to work on the PROGNOST XP device NOTE The operating personnel must be familiar with all...

Страница 14: ...1 5 Explosion Protection The PROGNOST XP series is not designated for use within areas with explosive hazards 2 1 6 Interaction with Other Devices Interactions with other devices are not known 2 1 7 E...

Страница 15: ...ices should be observed to ensure that they are working properly NOTE The properties of this device determined by emissions allow its use in industrial areas and in hospitals CISPR 11 class A When use...

Страница 16: ...ST XP The PROGNOST XP does not have an emergency stop switch 3 2 2 Emergency Stop Switch PROGNOST XPE PROGNOST XPE Akku The PROGNOST XPE and PROGNOST XPE Akku has the following emergency stop switch w...

Страница 17: ...brake bracket with the foot the brakes of the tabletop are released then the table top can be moved floating by hand 3 3 1 3 Hand Grips optional A long handle for the back of the table top and 2 hand...

Страница 18: ...ake rollers on the front side which must be locked with the arrester on the left and right side before a patient gets on or off the table 3 3 2 2 Brake Bracket for the Table Top By operating the brake...

Страница 19: ...accordingly 3 3 2 5 Deflector optional The deflectors prevent the U rail from touching the detector housing and thus facilitate the positioning of the table top above the image receiver 3 3 2 6 Latera...

Страница 20: ...light green Energy saving mode o Operational readiness is activated by pressing the release key PROGNOST XPE Akku green Device is ready for operation PROGNOST XPE PROGNOST XPE Akku orange Battery almo...

Страница 21: ...possible Replace or charge battery 5 7x The control has detected that the upper or lower limit switch is out of function Take the device out of order and notify PROTEC authorized service 8x Height di...

Страница 22: ...of order and the service authorized by PROTEC must be notified 3 3 3 2 Acoustic and Visual Status Indications at Column Drive Blockage In the event of a blocked column drive a visual indicator lamp st...

Страница 23: ...button 6 movement release button 6 Movement release button 7 Move the table top downwards slowly in combination with button 6 movement release button 8 Move the table top upwards slowly in combinatio...

Страница 24: ...ked positions with the brakes applied and with the supervision or assistance of the examiner as otherwise there is a risk of tipping over The patient must be given assistance when climbing on Position...

Страница 25: ...tton Store button see chapter 3 of the hand control must be pressed three times Afterwards use the synchronous adjustment button see chapter 3 to move the table top down until both lifting columns aut...

Страница 26: ...ioning the patient during the examination The patient may only be ascended and descended at the marked positions with the brakes applied and with the supervision or assistance of the examiner as other...

Страница 27: ...batter box is charged with the battery box charging station CAUTION The batter boxes may only be charged with the PROTEC approved charger The 20A fuse protecting the 12VDC voltage in the PROGNOST XPE...

Страница 28: ...ight to be adjusted until the discharge safety limit of the battery box is reached If this discharge safety limit is reached further movements are prevented to protect the battery box against deep dis...

Страница 29: ...ox 5 Wall holder 6 Charging socket 4 2 3 5 Charging the Battery Box The charging station is used to store and hold the battery box to be charged The wall holder is used for space saving wall mounting...

Страница 30: ...s point the battery capacity has reached 80 95 of the total capacity LED green Trickle charge mode Standby The charger is in trickle charge mode If the battery box remains connected to the charger aft...

Страница 31: ...in the image is insufficient In this case please contact PROTEC service department or a service company authorized by them immediately 5 3 Cleaning and Disinfection NOTE Caution Possible material cha...

Страница 32: ...PROGNOST XPE Akku Verify that the brake rollers are also braked if the arrester is in brake position Check that the table top brake moves smoothly if the brake bracket is applied Check the table top b...

Страница 33: ...in the event of improper use are also excluded Warranty work may only be carried out by trained specialists 5 4 5 Product Service Life All products of the PROGNOST XP series are designed for a service...

Страница 34: ...ection cable 3 x 1 5mm 16 AWG max length 5m Version 2 Mains voltage 115V 60Hz Mains connection cable 3 x 1 5mm 16 AWG Type SJT or better Max length 5m with Hospital Grade USA power plug The power supp...

Страница 35: ...devices will be influenced RF emissions CISPR 11 Class A The device is suitable for use in facilities other than residential areas and those that are directly connected to the public supply network w...

Страница 36: ...5 90 135 180 225 270 and 315 0 UT for 1 Period 70 UT for 25 30 Periods 0 UT for 250 300 Periods The quality of the supply voltage should correspond to that of a typical business or hospital environmen...

Страница 37: ...0 FM 5 kHz Deviation 1 kHz Sine 28 710 745 780 704 787 LTE Band 13 17 Pulse modulation 217 Hz 9 810 870 930 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18 Hz 28 1720 18...

Страница 38: ...ble top 250 kg carbon fiber table top Total weight without patient 103 kg Table height 760 mm Table top transverse displacement from center position 100 mm Table top longitudinal displacement from cen...

Страница 39: ...ar opening between the columns under the table top 1121 mm The brakes of the table top and wheels are mechanically operated 7 1 3 Dimensions PROGNOST XPE Akku Table top dimensions 200 cm x 75 5 cm or...

Страница 40: ...0 6 and 0 8 mm Al according to 21CFR 1020 30 m at 100 kV and a first half value layer thickness of 3 6 mm Al 7 2 1 Protection Type and Protection Class The PROGNOST XP has no protection class whereas...

Страница 41: ...Abbreviations 8 1 Symbols Atmospheric pressure limit Temperature limit Humidity limit Keep dry Fragile handle with care This way up Attention observe accompanying documents Refer to Instructions for...

Страница 42: ...on our website Disposal instructions WEEE Waste of Electrical and Electronic Equipment Protective earthing Do not exceed the maximum indicated weight Do not exceed the maximum indicated weight Tableto...

Страница 43: ...V1 0_2021 05 17 PROGNOST XP XPE XPE Akku Instructions for use 5038 0 0002 43 of 51 8 2 Type Label 8 2 1 Type Label PROGNOST XP 8 2 2 Type Label PROGNOST XPE 8 2 3 Type Label PROGNOST XPE Akku...

Страница 44: ...er Table top composite fiber 200cm Table top composite fiber 226cm Labels on the top of the table top Caution Watch out for possible crushing hazards to fingers or hands while moving the tabletop tabl...

Страница 45: ...ent table tops Table top carbon fiber Table top composite fiber 200cm Table top composite fiber 226cm Labels on the top of the table top Caution Watch out for possible crushing hazards to fingers or h...

Страница 46: ...agency in the USA Label lock release brake rollers Label next to the power input Warning label on the lower right cover Power supply version 8 3 3 Labels PROGNOST XPE Akku Labels on the front side of...

Страница 47: ...etop Carbon fiber tabletop Labels on the lower table top frame Caution Table use for patient positioning in radiographic rooms only UL Organization in the USA for the approval of electrical products F...

Страница 48: ...station Label for the positive pole of the 12VDC battery supply inside the supply station and on the battery box Label for the negative pole of the 12VDC battery supply inside the supply station and...

Страница 49: ...V1 0_2021 05 17 PROGNOST XP XPE XPE Akku Instructions for use 5038 0 0002 49 of 51 8 4 2 Positions of the Signs and Labels PROGNOST XPE...

Страница 50: ...V1 0_2021 05 17 PROGNOST XP XPE XPE Akku Instructions for use 5038 0 0002 50 of 51 8 4 3 Positions of the Signs and Labels PROGNOST XPE Akku...

Страница 51: ...ions for use 5038 0 0002 51 of 51 8 5 Abbreviations Mm Millimetres cm Centimetres lb Pound Kg Kilogram C Degree Celsius hPa Hectopascal DIN German Industry Standard EN European Standard CE CE marking...