Содержание 4486-9-1000L

Страница 1: ...hone 49 0 7062 92 55 0 E mail protec protec med com Version 2 0 Effective from 2020 06 30 Subject to alterations RAPIXX 2530V1V Wifi Set RAPIXX 2530V2V Wifi Set Digital X ray detector system Model ID...

Страница 2: ...stenfeld Comments and questions about the documentation please contact PROTEC GmbH Co KG In den Dorfwiesen 14 71720 Oberstenfeld Germany Tel 49 0 7062 92 55 0 Fax 49 0 7062 92 55 60 E Mail protec prot...

Страница 3: ...stallation 10 4 1 Check list 10 4 2 Connection of the components 10 4 3 Setup the network 10 4 4 W LAN connection to the Detector 11 4 5 Copying detector specific data 11 4 6 Installation of CONAXX 2...

Страница 4: ...2 Safety related controls 18 9 3 3 Maintenance 18 10 Liability 18 11 Guarantee Fehler Textmarke nicht definiert 12 Disposal 19 13 Electromagnetic compatibility EMC according to EN 60601 1 2 19 13 1 Ge...

Страница 5: ...pment as of the date of manufacture Revisions to the equipment subsequent to the date of manufacture will be addressed in service notes distributed by PROTEC GmbH Co KG Document Version Revision Date...

Страница 6: ...persons responsible for the application must have the specific knowledge according to legal requirements and regulations and must establish safe exposure procedures for these kind of systems Those pe...

Страница 7: ...D CONAXX 2 CD 1 3 System requirements For an optimal use of the DR detector it is necessary that the hardware fulfils the system requirements described in the document System requirements 1 4 Environm...

Страница 8: ...ray detector systems of the RAPIXX series are only intended for the usage by professional users which are specially trained on the operation of diagnostic X ray units according to the correspondent n...

Страница 9: ...lection of the used anti scatter grid 3 2 Notes for installing a new digital X ray system The generator cut off dose should be in the range between 3 0 and 3 3 Gy 3 3 Notes for upgrading an existing C...

Страница 10: ...After starting the tool all detected network interface cards and wireless adapters are listed Select the device with which the detector connection is to be established and select the entry INSTALL UN...

Страница 11: ...tically connect to the detector e g after battery changing 4 5 Copying detector specific data Install from the Detector CD the calibration data Normally the installation starts automatically If not pl...

Страница 12: ...he window Click on OPEN LINK in the ACQUISITION menu Start the Gain calibration by clicking the button GAIN CALIBRATION in the ACQUISITION menu The detector now acquires 8 images without x ray radiati...

Страница 13: ...ESULTS which is delivered with the detector The value is placed in the section SENSITIVITY in the field MAX COUNTS Depending on how the detector is built in it might be necessary to rotate the acquire...

Страница 14: ...ing please login to the dealer backroom and navigate to the page LICENSING Select the desired license and upload the hardware key with the function SET Download the license afterwards with GET For the...

Страница 15: ...close the toolbox If no error message appears at CONAXX 2 startup the connection is established correctly and the system is ready for use If no connection can be establish proceed as described in 7 3...

Страница 16: ...during operation with the modality Should an error appear during the daily operation with the modality e g communication error power blackout the error message will show hints to solve the problem Is...

Страница 17: ...e used such as Sani Cloth Plus a hospital grade EPA registered low to intermediate level product for hard non porous surfaces and equipment Use disinfectants in accordance with the manufacturer s inst...

Страница 18: ...must be completed every 12 months by the PROTEC service department or a PROTEC authorized service provider in order to ensure the reliable operation of the unit As the manufacturer PROTEC is responsi...

Страница 19: ...nd prevent any potential negative effects on human health and the environment which could otherwise arise from inappropriate waste handling In cases of any doubt contact PROTEC 13 Electromagnetic comp...

Страница 20: ...according to EN 60601 1 Type B Electrostatic sensitive device Large focus Small focus System CONAXX 2 Modality CONAXX 2 X ray generator Electrical product Do not dispose in the household waste 14 2 A...