PRS-812 Resistance Meter
[2] RESISTANCE RANGE
SELECTION
Two Triangular Arrows Buttons, UP (
↑
) and DOWN (
↓
), select Resistance
Measurement Range in single orders of magnitude while in MANUAL or
AUTOMATIC/MANUAL
modes
[3] TEST VOLTS
Manual Selection of <10, 10 or 100 volts in
MANUAL
selects initial test
voltage.
MANUAL
Test Voltage & Resistance Limits are as follows:
@<10V:
0.1 to <1.0E+05Ω (0.1Ω - <100KΩ)
@ 10V:
1.0E+03 to <1.0E+09Ω (1KΩ - <1GΩ)
@100V:
2.0E+05 to <2.0E+14Ω (200KΩ - <200TΩ)
NOTE:
Optimal minium measurement in
MANUAL
using 100
volts is 2.0E+05.
[4] RECORD/RECALL
If
REC
is not displayed in the lower left corner of the LCD, pressing
RECORD/
RECALL
once turns Memory Register ON [
REC
will then be displayed in
LCD]. If
REC
is ON and measurement data is stored in memory, pressing
RECORD/RECALL
once will provide access to the memory register. Pressing
RECORD/RECALL
successively will calculate and sequentially display
Minimum [
MIN
], Maximum [
MAX
] and Average [
AVG
] of data stored in the
Memory Register. If memory is
ON
and register contains stored data pressing
RECALL
will provide the following information:
First Press RECALL:
Provides access to data in Memory
NOTE: MEM
XX Flashes in LCD indicating number of data points in the
register and the last measurement is displayed. Pressing DOWN
(↓)
and UP
(↑)
displays other data points and their respective position in the register is
shown in the
MEM
XX section of the LCD.
Second Press RECALL:
Displays MIN Data Point in Memory
Third Press RECALL:
Displays MAX Data Point in Memory
NOTE:
When displayed,
OL
[Over Level] indicates a measurement greater
than (>) 2.0x10
14
ohms, which is beyond the measurement capability of the
PRS-812.
Fourth Press RECALL:
Displays
AVG
of all Memory Data Points
NOTE:
OL
[Over Level] measurements [>2.0x10
14
ohms] are not included in
displayed average (
AVG
) calculation.
Fifth Press RECALL:
Returns System to normal operations
NOTE:
If in
RECALL
mode, pressing
RESET
will return the instrument to normal
measurement operations.