E. TESTING THE HIGH END OF THE WRIST STRAPS “PASS” RANGE
Confi rm the toggle posi on is in the “Wrist Straps” posi on. Set the Variable Resistance Refer-
ence to 9.0x10
6
ohms. Using the metal banana plug end of the second reference lead, press the
PMT-872A’s “Push to Test” bu on and observe the instrument’s indica on LED lights. The following
PMT-872A indica ons should occur within approximately 1 second:
1. Ini ally, the
HIGH FAIL
and LOW BATT. LED’s should light briefl y.
2. The
HIGH FAIL
LED goes off , the
PASS
LED comes on and the LOW BATT. LED remains on.
3. The
PASS
LED remains on and the LOW BATT. LED goes off .
4. The audible BEEP begins to sound indica ng an acceptable resistance range.
F. TESTING THE WRIST STRAPS “
HIGH FAIL
” POINT
Confi rm the toggle posi on is in the “Wrist Straps” posi-
on. Set the Variable Resistance Reference to 1.05x10
7
ohms. Using the metal banana plug end of the second
reference lead, press the PMT-872A’s “Push to Test” but-
ton and observe the instrument’s indica on LED lights.
The following PMT-872A indica ons should occur within
approximately 1 second:
1. Ini ally, the
HIGH FAIL
and LOW BATT. LED’s should
light briefl y.
2. The
HIGH FAIL
LED remains on and the LOW BATT. LED goes off .
G. TESTING THE FOOTWEAR “LOW FAIL” POINT
Posi on the toggle switch
DOWN
, to the “Footwear: posi-
on, and set the Variable Resistance Reference to 7.75x10
5
Ohms. Using the metal banana plug end of the second
reference lead, press the PMT-872A’s “Push to Test” but-
ton and observe the instrument’s indica on LED lights.
The following three indica ons should take place within
approximately 1 second:
1. The
PASS
LED and the LOW BATT. LED come on.
2. The
PASS
LED goes off , the
LOW FAIL
LED comes on
and the LOW BATT. LED remains on.
3. The
LOW FAIL
LED remains on and the LOW BATT. LED goes off .
H. TESTING THE LOW END OF THE FOOTWEAR “PASS” RANGE
Set the Variable Resistance Reference to 1.1x10
6
ohms. Using the metal banana plug end of the
second reference lead, press the PMT-872A’s “Push to Test” bu on and observe the instrument’s
indica on LED lights. The following PMT-872A indica ons should occur within approximately 1
second: