16
4.
Point the unit away from surfaces and objects. Press and hold the yellow ZERO button until
the display indicates 0.00.
5.
With the operator’s fingers contacting the metal grounding snap on the back of the PCS-730B
Charging Source, press the + button switch (+1kV). Touch the tip of the charging wand directly
to the upper sensing plate of the CPM-720B Charge Plate Assembly and charge the CPM-720B
to approxi 1.25kV ±5%.
6.
Discharge the upper plate to ground, or short the upper and lower plates, until the PFM-711B
meter reads zero (0.00).
7.
With the operator’s fingers contacting the metal grounding snap on the back of the PCS-730B
Charging Source (Figure 7), press the - button switch (-1kV). Touch the tip of the charging
wand directly to the upper sensing plate of the CPM-720B Charge Plate Assembly and charge
the CPM-720B to approximately - 1.25kV ±5%.
8.
Discharge the upper plate to ground, or short the upper and lower plates, until the PFM-711B
meter reads zero (0.00).
Figure 7. Operator properly grounded to use the PCS-730B
Grounding Snap on the
back of the PCS-730B
Wrist Strap