11

OPERATING INSTRUCTIONS

These instructions are for general use. When required, refer to 

the detailed information in the second half of this manual.

l

   Ensure  that 

ONLY

 suitably trained clinicians operate  

the device.

l

   Ensure that the warnings and cautions are observed.

l

   Ensure that the patient exhibits no contra indications.

l

   Ensure the unit has been cleaned prior to first use  

(refer to page 12 for detailed cleaning guidance).

l

    Mains electrical power is not necessary if the battery is 

adequately charged.

l

    The device may be operated whilst connected to the mains 

electrical supply using the Power Adaptor provided.

l

   Ensure that the footswitch is connected.

l

   The reservoir should be removed prior to filling.

l

   Fill the reservoir with a maximum of 700ml of warm water at 

40°C. Monitor temperature regularly to ensure that patient 
comfort and safety is maintained. Refill as necessary.

l

   Fit a new Propulse QrX Single Use Tip to the QrX Handle.

l

   Set Pressure/Flow switch (6) to minimum (by turning  

anti-clockwise).

l

    Turn the Power Switch (7) to on (marked “I”).

l

    With the footswitch depressed, check that the power status 

(9) is adequate (green indicator). If the power status is 
inadequate, (amber or red indicator), the Propulse NG 
Electronic Ear Irrigation System must be connected to the 
mains electrical supply using the Power Adaptor provided. 

l

   Adjust the water pressure/flow to an appropriate value that 

is compatible with the treatment requirements and patient 
comfort.

l

   Direct the irrigator tip into the noots tank and switch on  

the machine for 10-20 seconds in order to circulate the 
water through the system and eliminate any trapped air or 
cold water. Ensure the water is warm before presentation to 
the patient.

l

    During treatment you can pause the flow by releasing the 

footswitch or firmly pressing and holding the water flow 
button (5) on the QrX Handle (3).

l

   After treatment empty the reservoir and operate the device 

to purge any residual water.

l

    Remove the Propulse QrX Tip and dispose of in accordance 

with local authority guidelines.

l

    Turn off Power Switch (7) after use and connect to the  

Power Adaptor to recharge the battery (if required).

l

   Clean the Propulse NG Electronic Ear Irrigation System  

unit every morning prior to use, with a cleaning tablet  
(See page 12).

l

   The Propulse NG Electronic Ear Irrigation System should 

only be transported in a Propulse approved carry case to 
prevent damage or contamination.

l

   If any changes in performance occur, turn off the Propulse 

NG Electronic Ear Irrigation System, disconnect from the 
mains electricity supply and  

DO NOT

 use. Please refer the machine to the Mirage 

Health Group Certified Service Centre (See page 14).

DETAILED INFORMATION

Fitting the Footswitch

The Propulse footswitch is connected to the main unit by 
a socket (10) in the side of the device. The Propulse NG 
Electronic Ear Irrigation System will 

NOT

 operate unless the 

footswitch is connected.

Filling the water container

It is recommended that:
i)    The water container/reservoir (1) is removed from the device 

for filling and that the lid is always in place when the water 
container is in-situ on the device.

ii)   The water reservoir must 

NOT

 be filled over the 700ml 

graduation mark to eliminate the risk of spillage.  
(The graduations on the reservoir are for indication only).

iii)  Water above 40°C should 

NOT

 be used in order to avoid 

the risk of scalding and burns to the user and patient.  
Water below 40°C should 

NOT

 be used in order to avoid 

the risk of patient discomfort and dizziness.

Fitting the Propulse QrX Tip

The Propulse NG Electronic Ear Irrigation System is supplied 
with Ten Propulse QrX Single Use Tips. Use One Propulse QrX 
Tip per treatment.

To fit a Propulse QrX Tip 
1)  Remove tip from packaging - Tips are non-sterile.
2)   Push Tip onto QrX Handle whilst retracting the  

QrX Locking Collar.

3)   Release the QrX Locking Collar to secure tip to the  

QrX Handle.

To remove a Propulse QrX Tip,
1)  Retract the QrX Locking Collar.
2)   Grasp the used QrX Tip between forefinger and thumb.  

Pull gently from the QrX Handle.

3)   Dispose of used tip in accordance with local authority 

guidelines. 

DO NOT

 reuse tips.

Содержание NG

Страница 1: ...User Manual English Language Version 0086 ...

Страница 2: ...R R Dressing Scissors INS0024 Propulse Tilley Style Aural Dressing Forceps INS0022 Propulse Cleaning Tablets x200 per pack CL0001 Propulse QrX Single Use Tip x100 per pack KIT4002 Propulse Carry Case KIT1005 Propulse Headlamp INS0014 Propulse Replacement Mushroom Valves Washers KIT1006 Replacement Reservoir Lid KIT4004 Instruments Professional Ear Care with Integrity Accessories Professional Ear C...

Страница 3: ... accessories with other devices l If the device is to be used for domestic visits it is strongly recommended that a Propulse Carry Case is used to prevent damage and contamination l The Propulse NG Electronic Ear Irrigation System is not user repairable and should be returned to your Propulse supplier or the Mirage Health Group Certified Service Centre UK customers only for service and or repair I...

Страница 4: ...otswitch 12 is connected to the main body via a jack plug socket connection 10 The device will only function if the footswitch is connected The footswitch controls the flow of water Depressing the footswitch starts the flow releasing the footswitch stops the flow The flow pressure can be adjusted by rotating the pressure flow control knob 6 on the machine Please note Residual water in the handle a...

Страница 5: ... shielding The power frequency magnetic field should be measured in the intended installation location to assure that it is sufficiently low Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80 MHz to 2 5 GHz 3 Vrms 3 V m Portable and mobile RF communications equipment should be used no closer to any part of the Propulse NG including cables than the recommended se...

Страница 6: ...ove the Propulse QrX Tip and dispose of in accordance with local authority guidelines l Turn off Power Switch 7 after use and connect to the Power Adaptor to recharge the battery if required l Clean the Propulse NG Electronic Ear Irrigation System unit every morning prior to use with a cleaning tablet See page 12 l The Propulse NG Electronic Ear Irrigation System should only be transported in a Pr...

Страница 7: ...stem for a few seconds to fill the pump and hose Propulse QrX Tips are available in boxes of 100 individually wrapped non sterile tips from your normal Propulse supplier or from Mirage directly UK only Propulse QrX Tips are clearly branded with the Propulse logo on the tip and it s packaging Only branded Propulse QrX Tips should be used with the Propulse NG Electronic Ear Irrigation System Replaci...

Страница 8: ...el 0845 130 5445 The clinical procedures relating to the use of ear irrigators can be found on the following websites www earcarecentre com www entnursing com earcare htm Mirage is not responsible for the content or maintenance of third party internet sites Mirage can also advise on the availability of training courses offered by the relevant organisations Warranty The Propulse NG Electronic Ear I...

Страница 9: ...td BioPark Hertfordshire Broadwater Road Welwyn Garden City Hertfordshire AL7 3AX UK Tel 0845 130 5440 Fax 0845 130 6440 Overseas Tel 44 845 130 5440 Fax 44 845 130 6440 www miragehealthgroup com uksales miragehealthgroup com internationalsales miragehealthgroup com 0086 Copyright Mirage Health Group Ltd August 2012 PPNG UK MANUAL MHG 0812 ...

Отзывы: