Guidance & manufacturer’s declaration - electromagnetic immunity

The Propulse NG is intended for use in the electromagnetic environment specified below. The customer or the user of the Propulse NG 

should assure that it is used in such an environment.

IMMUNITY test

IEC 60601 test level Compliance level

Electromagnetic environment - guidance

Electrostatic discharge (ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered with 
synthetic material, the relative humidity should be at least 30%.

Electrical fast
transient/burst
IEC 61000-4-4

± 2 kV for power supply lines
± 1 kV for input/output lines

± 2 kV for power supply lines
± 1 kV for input /output lines

Mains power quality should be that of a typical commercial or  
hospital environment.

Surge
IEC 61000-4-5

± 1 kV line(s) to line(s)
± 2 kV line(s) to earth

± 1 kV differential mode
± 2 kV common mode

Mains power quality should be that of a typical commercial or  
hospital environment

Voltage dips, short  
interruptions and
voltage variations
on power supply 
input lines
IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle
40 % U

(60 % dip in U

T

)

for 5 cycles
70 % U

(30 % dip in U

T

)

for 25 cycles
<5 % U

(>95 % dip in U

T

)

for 5 s

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle
40 % U

(60 % dip in U

T

)

for 5 cycles
70 % U

(30 % dip in U

T

)

for 25 cycles
<5 % U

(>95 % dip in U

T

)

for 5 s

Mains power quality should be that of a typical commercial or hospital 
environment. If the user of the Propulse NG requires continued operation 
during power mains interruptions, it is recommended that the Propulse NG be 
powered from an uninterruptible power supply or the internal battery.

Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8

3 A/m

3 A/m

If there is a reduction in delivery performance  it may be necessary to position 
Propulse NG further from sources of power frequency magnetic fields or to 
install magnetic shielding. The power frequency magnetic field should be 
measured in the intended installation location to assure that it is sufficiently low.

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

3 Vrms
150 kHz to 80 MHz

3 V/m
80 MHz to 2,5 GHz

3 Vrms

3 V/m

Portable and mobile RF communications equipment should be used no closer 
to any part of the Propulse NG, including cables, than the recommended 
separation distance calculated from the equation applicable to the frequency of 
the transmitter.
Recommended separation distance:
d = 1,2 

√ 

P

d = 1,2 

√ 

P 80 MHz to 800 MHz

d = 2,3 

√ 

P 800 MHz to 2,5 GHz

where P is the maximum output power rating of the transmitter in watts 
(W) according to the transmitter manufacturer and d is the recommended 
separation distance in metres (m). Field strengths from fixed RF transmitters, 
as determined by an electromagnetic site survey, (a) should be less than the 
compliance level in each frequency range (b). Interference may occur in the 
vicinity of equipment marked with the following symbol:

Note 1:

 UT is the a.c. mains voltage prior to application of the test level.

Note 2:

 At 80 MHz and 800 MHz, the higher frequency range applies.

Note 3:

 These guidelines may not apply in all situations. Electromagnetic 

propagation is affected by absorption and reflection from structures, objects and 

people.

(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/

cordless) telephones and land mobile radios, amateur radio, AM and FM radio 

broadcast and TV broadcast cannot be predicted theoretically with accuracy. 

  To assess the electromagnetic environment due to fixed RF transmitters, an 

electromagnetic site survey should be considered. If the measured field strength 

in the location in which the Propulse NG is used exceeds the applicable RF 

compliance level above, the Propulse NG should be observed to verify normal 

operation.  

  If abnormal performance is observed, additional measures may be necessary, 

such as re-orienting or relocating the Propulse NG.

(b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less 

than 3 V/m.

Содержание NG

Страница 1: ...User Manual English Language Version 0086 ...

Страница 2: ...R R Dressing Scissors INS0024 Propulse Tilley Style Aural Dressing Forceps INS0022 Propulse Cleaning Tablets x200 per pack CL0001 Propulse QrX Single Use Tip x100 per pack KIT4002 Propulse Carry Case KIT1005 Propulse Headlamp INS0014 Propulse Replacement Mushroom Valves Washers KIT1006 Replacement Reservoir Lid KIT4004 Instruments Professional Ear Care with Integrity Accessories Professional Ear C...

Страница 3: ... accessories with other devices l If the device is to be used for domestic visits it is strongly recommended that a Propulse Carry Case is used to prevent damage and contamination l The Propulse NG Electronic Ear Irrigation System is not user repairable and should be returned to your Propulse supplier or the Mirage Health Group Certified Service Centre UK customers only for service and or repair I...

Страница 4: ...otswitch 12 is connected to the main body via a jack plug socket connection 10 The device will only function if the footswitch is connected The footswitch controls the flow of water Depressing the footswitch starts the flow releasing the footswitch stops the flow The flow pressure can be adjusted by rotating the pressure flow control knob 6 on the machine Please note Residual water in the handle a...

Страница 5: ... shielding The power frequency magnetic field should be measured in the intended installation location to assure that it is sufficiently low Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80 MHz to 2 5 GHz 3 Vrms 3 V m Portable and mobile RF communications equipment should be used no closer to any part of the Propulse NG including cables than the recommended se...

Страница 6: ...ove the Propulse QrX Tip and dispose of in accordance with local authority guidelines l Turn off Power Switch 7 after use and connect to the Power Adaptor to recharge the battery if required l Clean the Propulse NG Electronic Ear Irrigation System unit every morning prior to use with a cleaning tablet See page 12 l The Propulse NG Electronic Ear Irrigation System should only be transported in a Pr...

Страница 7: ...stem for a few seconds to fill the pump and hose Propulse QrX Tips are available in boxes of 100 individually wrapped non sterile tips from your normal Propulse supplier or from Mirage directly UK only Propulse QrX Tips are clearly branded with the Propulse logo on the tip and it s packaging Only branded Propulse QrX Tips should be used with the Propulse NG Electronic Ear Irrigation System Replaci...

Страница 8: ...el 0845 130 5445 The clinical procedures relating to the use of ear irrigators can be found on the following websites www earcarecentre com www entnursing com earcare htm Mirage is not responsible for the content or maintenance of third party internet sites Mirage can also advise on the availability of training courses offered by the relevant organisations Warranty The Propulse NG Electronic Ear I...

Страница 9: ...td BioPark Hertfordshire Broadwater Road Welwyn Garden City Hertfordshire AL7 3AX UK Tel 0845 130 5440 Fax 0845 130 6440 Overseas Tel 44 845 130 5440 Fax 44 845 130 6440 www miragehealthgroup com uksales miragehealthgroup com internationalsales miragehealthgroup com 0086 Copyright Mirage Health Group Ltd August 2012 PPNG UK MANUAL MHG 0812 ...

Отзывы: