![ProMinent DULCOTEST CLO 1 Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/prominent/dulcotest-clo-1/dulcotest-clo-1_operating-manual_1616853037.webp)
11.6 Temperature and climate
Range
Applications
Value
Storage temperature
-5 - 50 ℃
Ambient temperature
5 - 50 ℃
Measuring temperature
CLO 1 swimming pool
5 - 45 ℃
CLO 2 hot water
40 - 70 ℃
Air humidity
95 % relative air humidity
non-condensing
11.7 Rated gradient, drift and response time
Rated gradients
Sensor
Rated gradient *
CLO 1/2 - 2 ppm
6 mA/ppm
CLO 1/2 - 10 ppm
1.2 mA/ppm
* at pH = 7.2; CLO 1 T = 30 °C; CLO 2 T = 50 °C; Q = 60 l/h in DGMa
Gradient spread: rated gradient ± 50 %
Drift: < 5 % / months, depending on water quality and temperature
Response time
Value
T90(up)
approx. 60 seconds
T90(down)
approx. 60 seconds
Technical Specifications
37