6 Commissioning
n
Users' qualifications: Technical expert, see
Chapter 1.2 “Users' qualifications ” on page 10
CAUTION!
Incorrect metering due to sensor failure
Possible consequence: Slight or minor injuries, material damage
–
If a sensor fails then there may be an incorrect reading at the input of the controller/measuring
device
–
This may result in uncontrolled metering
–
The operator must therefore ensure that no subsequent damages can result from this
CAUTION!
–
Do not switch off the measuring system during intermittent operation!
–
A run-in time will be necessary following operation without chlorine. Switch on metering
equipment with a time delay if necessary!
–
If no chlorine is added for a long time ( ~ 1 week) then the sensor must be reactivated using
emery. The sensor is once more ready for use after expiration of the run-in time and a repeat
calibration.
Run-in time
The sensor requires a specific run-in time to display a steady display value.
Initial commissioning:
1 - 24 h (typically 6 h)*
After cleaning the working electrode and counter
electrode:
1 - 24 h (typically 6 h)*
Repeat operation:
1 - 24 h (typically 6 h)*
Electrolyte replacement:
1 - 3 h
* the exact run-in time depends on the application.
Commissioning
22