Calibration of zero point (zp)
Complete Operating Menu / Description
D1C2-cond.-039-GB
cali. zero point ?
PAR1
cali. zero point
probe in air!
accept value?
12
µ
S/cm
please wait!
To determine the precise zero point (zp) of the sensor, hold the sensor in the air and select the second
menu option. The DULCOMETER
®
D1C displays the conductivity without zero point correction. Press
Enter when the value is constant.
The setting menu is inactive when “PAR2“ appears.
During the calibration, the metering is reduced to the set basic load and control stops. The limit value
monitor and the fault diagnosis system are reset. The standard signal of the “measured value“ output is
frozen.
Cell constant calibration (cc)
D1C2-cond.-043-GB
calibration cc
probe in buffer!
accept value?
1538
µ
S/cm 8.54
calibration cc ?
PAR1
calibration cc
cc: 8.54 /cm
LF: 100
µ
S/cm
calibration cc
cc: 8.50 /cm
please wait!
To determine the precise cell constant (cc) of the sensor, place the sensor in a calibration solution
(sample 1) with a known conductivity and select the second menu option. The DULCOMETER
®
D1C
displays the conductivity which it has calculated with the current parameters. Press Enter when the value
is constant.
Select the next menu option and enter the conductivity of the calibration solution (arrow keys). Press Enter.
The D1C displays the recalculated cell constant and saves the value.
The setting menu is inactive when “PAR2“ appears.
During the calibration, the metering is reduced to the set basic load and control stops. The limit value
monitor and the fault diagnosis system are reset. The standard signal of the “measured value“ output is
frozen.
Possible values
Initial value
Increment
Lower value
Upper value
Remarks
Conductivity
Measured value
0.1
µ
S/cm
0
µ
S/cm
200
µ
S/cm
Measuring range 200
µ
S/cm
Calibration
1
µ
S/cm
0
µ
S/cm
2000
µ
S/cm
Measuring range 2000
µ
S/cm
solution (LF)
0.01 mS/cm
0
µ
S/cm
20 mS/cm
Measuring range 20 mS/cm
0.1 mS/cm
0
µ
S/cm
200 mS/cm
Measuring range 200 mS/cm
1 mS/cm
0
µ
S/cm
2000 mS/cm
Measuring range 2000
µ
S/cm
BA_DM_138_09_06_GB.p65
21.09.2006, 11:22 Uhr
15