12
General settings
access c.:
operating menu:
5000
restricted
english
=
=
D1C2-cond.-016-GB
Access to all setting menus can be blocked with an access code !
general setting
information
ident-code: D1CA
software version
DxL6xxxxxxxxx
D1C-L0 FW-2.03
alarm relay
active
input HW:
input SW:
rev. y
xx.xx.xx
Restricted Operating Menu / Description
Possible values
Initial value
Increment
Lower value
Upper value
Remarks
Setpoint
100
µ
S/cm
0,1
µ
S/cm
-10
µ
S/cm
210
µ
S/cm
Measuring range 200
µ
S/cm
1000
µ
S/cm
1
µ
S/cm
-100
µ
S/cm
2100
µ
S/cm
Measuring range 2000
µ
S/cm
10 mS/cm
0,01 mS/cm
-1 mS/cm
21 mS/cm
Measuring range 20 mS/cm
100 mS/cm
0,1 mS/cm
-10 mS/cm
210 mS/cm
Measuring range 200 mS/cm
1000 mS/cm
1 mS/cm
-100 mS/cm
2100 mS/cm
Measuring range 2000 mS/cm
Control with dead zone
2 setpoints necessary
Setpoint 2
≥
Setpoint 1
Adjustment of measuring
range on page 9/14
Control parameter xp
10 %
1 %
1 %
500 %
xp referred to measuring range
Control parameter Ti
off
1 s
1 s
9999 s
Function off = 0 s
Control parameter Td
off
1 s
1 s
2500 s
Function off = 0 s
Manual metering
0 %
1 %
-100 %
+100 %
Possible values
Initial value
Increment
Lower value
Upper value
Remarks
Alarm relay
active
active
not active
Access code
5000
1
1
9999
Language
as per
as per
identity code
identity code
Operating menu
restricted
restricted
complete
Abbreviations for control variables:
xp: 100 %/Kp (inverse proportional coefficient)
T
i
: Integration time of I-controller (s)
T
d
: Differential time of D-controller (s)
BA_DM_138_09_06_GB.p65