About setpoint and tolerances
P
P
P
P
P
P
P
B0230
calibration
stroke feedback
volume (stroke)
0.347 mL
stroke feedback
set-point
volume (stroke)
0.347 mL
lower tolerance
-010.00 %
tolerance
tolerance (%)
upper tolerance
+020.00 %
Specify the permitted range for the stroke volume V/H using the setpoint of the stroke volume V/H
and tolerances in %:
1.
Follow the
‘calibration
è
stroke feedback
è
setpoint’ menu path and press [P].
ð
The
‘volume (stroke)’ menu item displays the setpoint currently saved.
2.
Start the metering pump.
ð
The current measured value is displayed.
3.
Turn the stroke length adjustment knob until the required setpoint is displayed.
4.
Press
[P].
ð
The measured value displayed is saved as a setpoint and the
‘tolerance tolerances’
menu item appears.
The setpoint is valid as 100 %.
5.
Stop the metering pump.
6.
Press
[P].
7.
Use the
[arrow keys] to enter the ‘lower tolerance’ and press [P].
8.
Use the
[arrow keys] to enter the ‘upper tolerance’ and press [P].
ð
The continuous display appears again.
The required setpoint can also be entered without the need to use the stroke adjustment dial
and the pump to be running, by using the [arrow keys] directly in ‘volume (stroke)’ .
Set up
43