calibration
zero p.:
take over value?
4.00 mA
calibration
DPD-value
1.55 ppm
calibration
zero p.: 4.00 mA
slope
6.50 mA/ppm
calibration
zero p.: 4.00 mA
slope
6.50 mA/ppm
calibration
Zeropoint
4.00 mA
calibration
DPD-value
01.55 ppm
A0018_GB
14.00 mA
Fig. 9: Calibration of All Amperometric Measured Variables
Fault message
Condition
Remarks *
Calibration not possible!
Slope too low
Slope too low
(< 20 % of standard slope)
Repeat calibration
Calibration not possible!
Slope too high
Slope too high
(> 300 % of standard slope)
Repeat calibration
DPD value too low
DPD > x.xx ppm
DPD < 2 % of measuring range
Repeat calibration after addition of
metering medium or fit sensor
suitable for the process
Calibration not possible!
Zero point low
< 3 mA
(only with 4 - 20 mA sensors)
Check sensor/cable
Repeat calibration in water without
metering medium
Calibration not possible!
Zero point high
> 5 mA
> 6 mA for 0.5 ppm chlorite
Check sensor/cable
Repeat calibration in water without
metering medium
* Please also note the operation manual for the respective sensor
Only the slope can be calibrated in the restricted operating menu of
the DULCOMETER
®
D1Cb / D1Cc.
Both the zero point and the slope can be calibrated in the complete
operating menu of the DULCOMETER
®
D1Cb / D1Cc.
CAUTION!
Correct sensor operation / Run-in time
Damage to the product or its surroundings
– Correct measuring and dosing is only possible if
the sensor is working perfectly
– Please read the operating manual for the sensor
– Please also read the operating manuals for the fit‐
tings and other components used
– It is imperative that the run in periods of the sen‐
sors are adhered to
– The run in periods should be allowed for when
planning commissioning
– It may take a whole working day to run-in the
sensor
Testing the sensor
Start up
25