4.
Plug the mains plug into the socket and switch on the mains
voltage
6.1
Priming and bleeding
Start the control process, if correctly installed, the lines bleed them‐
selves.
6.2
Adjust the flow sensor switching point
Sample water may escape.
1.
Adjust the flow to 50 l/h using the ball valve
2.
Hold the flow sensor in place and loosen the mounting clip a
little
3.
Press the float to 40 l/h using the flow sensor
ð
The error message should disappear.
4.
Hold the flow sensor in this position and tighten the mounting
clip
5.
Then re-adjust the flow required using the ball valve
6.
Acknowledge any error message that occurs
7.
Reset any possible consequences of this in the overall instal‐
lation
8.
Check the threaded connector for leak-tightness
6.3
Calibration
You have to regularly calibrate the sensor during ope‐
ration. That means: 24 hours after initial calibration
and then weekly thereafter.
Please take into account any differing national regula‐
tions.
The flow sensor should switch when
the flow falls (flow sensor is con‐
nected as an NC).
Start up
23