Dulcodes UV-Desinfektionsanlage
5.3
Installing Lamp Protective Tubes
ATTENTION!
Examine for signs of damage before installing the lamp protective tubes!
Do not install damaged protective tubes!
Use supplied C-wrench to release the retaining fixtures for the protective tubes
Remove retaining fixtures for protective tubes by turning out of the radiation chamber
Fit supplied O-ring approx. 40 mm over the open end of the lamp protective tube
Insert lamp protective tube in radiation chamber
ATTENTION!
Make sure the protective tube is fitted correctly!
The lamp protective tube may protrude by a maximum of 40 mm and exhibit no angle
offset!
Check whether the Teflon ring is located in the groove provided in the retaining fixture of the
protective tube (see Fig. 2). Otherwise fit the Teflon ring into the retaining fixture of the
protective tube such that it snaps into the groove.
Slide the retaining fixture over the protective tube and screw into the radiation chamber;
tighten by hand.
Use supplied C-wrench to tighten the protective tube retaining fixtures; very little force is
required for this purpose
5.4
Installing and Connecting UV Lamps
WARNING!
• Switch off the master switch or disconnect the power plug before installation and
connecting the UV lamps!
• Place the UV lamps into operation only in their installed state!
UV-C radiation is harmful to the eyes and skin!
Install the UV disinfection system in accordance with regulations and instructions before
starting up the UV lamps!
• Do not change/modify the lamp connection cables already fitted!
• Do not change/modify the distance between the connector and lamp cover!
Otherwise it will not be possible to ensure that the lamp rests against the closed end of
the protective tube. This position, however, is the prerequisite for safe and reliable
disinfection!
ATTENTION!
Do not touch the glass of the UV lamps with bare hands!
Fingerprints burn into the glass and can cause premature failure. Before fitting the lamps,
remove fingerprints with a cloth moistened with alcohol!
Systems without connector on lamp cover
Fit O-ring in the groove provided on the protective tube retaining fixture
Insert UV lamp in the protective tube so that it protrudes by approx. 100 mm
Plug in connector on UV lamp
The connector can be plugged in only in certain positions, it may be necessary to turn it
through 90
before attempting to plug in again
Completely insert the UV lamp into the protective tube
Secure UV lamp cover on retaining fixture for protective tube with knurled nut; only slight force
is required
Systems with connector on UV lamp cover
Fit O-ring in the groove provided on the protective tube retaining fixture
Plug in connector on UV lamp
The connector can be plugged in only in certain positions, it may be necessary to turn it
through 90
before attempting to plug in again
Insert UV lamp into protective tube until the UV lamp cover comes to rest against the retaining
fixture for the protective tube
Secure UV lamp cover on retaining fixture for protective tube with knurled nut. Use only very
little force to tighten the knurled nut.
Plug in circular connector at the socket on the UV lamp cover and secure with knurled nut.
Use only very little force to tighten the knurled nut.
Assembly and Installation