Function
Options
Info
Addr
1 ... 255
Selection of the device
address
MBUS
ASCII
RTU
Selection ASCII or RTU
6.4.2 Configuration of fault current (ERLV)
A0625
Fig. 11: Configuration of fault current (ERLV)
The 4-20 mA signal output can be utilized in the event of a system error on the
DULCO
®
turb C in order to indicate the problem by activating the
[ERLV ON]
function. In doing so, fault currents of 4.00 mA, 2.00 mA or 0 mA can be
selected. In the event of an error, the configured current value is output to the
controller, regardless of which measured value is pending. If the function is
switched off by setting
[OFF] then the 4-20 mA signal output remains unaffected
by fault conditions.
Commissioning
37