AMMONIA User Manual
AMMONIA_User 9/09
5
7
5.8 System Diagnostic
To view the Analyzer’s
Diagnostic
press
ENTER
at the power up
or day-time display
.
Key
ENTER
@
Diagnostic
.
Displays the state of the Relay 1 to Relay 5 AC power fuse.
Displays ‘
OPEN
’ if AC fuse has failed.
The
Internal 2.5V
level corrects analog voltages measured
by the Analyzer. Displays from 2.4500 to 2.5500.
The
15VDC Power
level is the unregulated voltage @ the
Analyzers
+DC Power Output
terminals. Displays from 15
to 23 Volts depending on AC line voltage
and load on the
+DC Power Output
supply.
Displays the time and date of the last Analyzer reset
or the time and date of the most recent
Load Factory
was executed.
An
Admin Password
@
Default
has not been modified
from ‘
AAAA
’. If modified displays ‘
Changed
’.
An
Watchdog Resets
should always display
0
.
An increasing number of
Resets
indicates corrupted
firmware or Analyzer electrical fault or interference.
The Analyzer
Firmware Version
indicates the date of
issue of the software operating the Analyzer.
Thu 16:54:10
S/N: A076X0486
Diagnostic
Communicate
Relay Fuse
OK
Internal 2.5V
2.4743
15VDC Power
16.97
Reset to Factory
10:32 20/08/07
Admin Password
Default
Watchdog Resets
0
Firmware Version
A587
Sidebar:
Reset to Factory:
Sets all volume-water meter totals for this year and the previous year to zero.