AMMONIA User Manual
AMMONIA_User 9/09
3
7
2.6 Sensor Diagnostics
To view sensor
Diagnostics
,
key
UP
or
DOWN
to the target sensor
and press
ENTER
.
Key
UP
and
ENTER
@
Diagnostic
.
Sensor Type
;
NH3
in this example.
Also displays pH, Temperature, unused, or 4-20mA Input
Current State
may also display
Alarmed
,
Fail Calibrate
,
4-20mA Open
(4-20mA Inputs only), or
Overrange
( Hardware fault )
Current value of the sensor.
With user set digits after the decimal and user set units.
Thermal Compensation
is used with pH and NH3
Gain Multiply
is the value required to make the displayed
NH3 value match your last calibration.
Default Gain
is the
Gain
after a
Factory Reset
.
An increasing
Gain Multiply
usually
Indicates a fouling sensor.
NH3
1.30 ppm
Diagnostic
Calibrate
Sensor Type
NH3
Current State
Operational
D
Displayed Value
1.30 ppm
Diagnostic
and
Compensation
Thermal Comp.
Gain Multiply
5.7146
Default Gain
5.6000
continued
Sidebar:
Diagnostic
displays how the sensor is configured, compensated and calibrated.
This is where you go if you have a non-obvious sensor problem.