USER’S MANUAL. OS-782
Page 35
02-2010
En
g
lis
h
4.16.3 XY mode
This mode is only applicable to
CH1
and
CH2
. With the
XY
mode selected,
CH1
is displayed on the horizontal axis and
CH2
is on the vertical axis; when the
oscilloscope is under the sampling mode in which no trigger is found, the data appear in
light spots.
Operations for various control keys are shown as below:
•
The
CH1 VOL
and
CH1 ZORE
for CH1 are used to set the horizontal
scale and position.
•
The
CH2 VOL
and
CH2 ZORE
for CH2 are used to set the vertical
scale and position continuously.
The following functions do not work in the XY display mode:
•
Reference or digital value waveform.
•
Cursor.
•
Auto Setting.
•
Time base control.
•
Trigger control.
4.17
Waveform Saving Setups
The oscilloscope can save 4 waveforms, which can be displayed on the screen
with the present waveform. The recalled waveform saved in the memory cannot be
adjusted.
The
waveform saving /recalling menu
is described in the following list:
Function menu
Setups
Description
Signal source
CH1
CH2
MATH
Select the displayed waveform which you want to
save.
Wave
A, B, C,
D
Select the address for saving or recalling a
waveform.
Saving
Store the waveform of a selected signal source into
the selected address.
Addresses A, B,
C, D
Close
Start
Close or start displaying the waveforms stored in
address A, B, C or D.