Installation & Owners Manual

Proma Dental Light

!

120V

60HZ

1A

1 AMP FUSES

INPUT VOLTAGE:

CURRENT:

111153

P/N 93155

CONFORMS TO
UL STD 60601-1

CERTIFIED TO

CAN/CSA STD C22.2 No. 601.1

WARNING:

REPLACE FUSE AS MARKED

RISK OF FIRE - CLASS 1

S

/N

P

L

T

A

50

00

00

1

A-5120 Ceiling Mounted Light

Specifications......................................   2

Operating Conditions..........................     2

Storage & Shipping Conditions...........    2   

Classifications.....................................     2

Technical Support................................2

Installation Overview...........................   2

Pre-Installation Notes..........................2
    Recommended Service Tools
    Electrical Connections and Supply
    Cautions on Installation and Use

Installation...........................................   3

Cleaning Instructions...........................4

Wiring Diagram & Warranty.................  4

Flex Arm Counterbalance
Adjustment..........................................     5

On   /   Off Switch Lever
Installation...........................................   5

Drill Template.......................................6

     Table of Content

     Serial Numbers

Please locate and write down the Serial 
Numbers  on  the  A-5120  Dental  Light 
and  Transformer.  Having  this  informa-
tion available will aid in the identification 
of your dental light and transformer.

Dental Light Serial Number

Transformer Serial Number

The Proma label is located in two loca-
tions  on  the  dental  light  system.  The 
dental light serial number is located on 
the light arm and the Transformer serial 
number  is  located  on  top  of  the  trans-
former housing

The serial number is located along the 
left edge of the label

     Bulb Replacement

Note:

 A  replacement  bulb  is  included. 

Please  locate  it  and  store  in  a  conve-
nient  place.  Order  replacement  Light 
Bulb Proma #93146

WARNING!

Turn the light off and allow cooling be-
fore removing the shield. Failure to do 
so can cause injury by burning. Always 
handle bulbs with tissue paper, do not 
allow skin to contact the bulb, as it will 
shorten the life.

Remove the shield by gently squeezing 
the ends and pull it toward you

Remove the bulb by wrapping in tissue 
and gently pulling forward

The replacement bulb should be 
pushed in gently. The lamp pins must 
be securely in the lamp socket

Replace the shield by squeezing the 
sides and gently pushing it into the 
housing. You will hear a “Click” when 
the shield seats properly.

WARNING!

Failure to properly seat the shield 
could allow it to fall off and cause injury

Turn  on  the  light  and  check  the  focus 
by  shinning  the  light  on  a  surface  27” 
away. The light pattern should be even 
with defined edges over an area  of 4” 
x  61⁄2”.  If  necessary,  rotate  the  focus 
adjustment.

Focus Adjustment Screw

730 E. Kingshill Place  •  Carson, CA 90746
Phone 310.327.0035  •  Fax 310.327.4601

Document 9050736 rev C 06/07