Proline Promag L 400
34
Hauser
Nominal diameter
Limit values for absolute pressure in [mbar] ([psi]) for fluid temperatures:
[mm]
[in]
+25 °C (+77 °F)
+90 °C (+194 °F)
200
8
200 (2.9)
290 (4.2)
250
10
330 (4.8)
400 (5.8)
300
12
400 (5.8)
500 (7.3)
Flow limit
The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The
optimum flow velocity is between 2 to 3 m/s (6.56 to 9.84 ft/s). Also match the velocity of flow (v)
to the physical properties of the fluid:
• v < 2 m/s (6.56 ft/s): for abrasive fluids (e.g. potter's clay, lime milk, ore slurry)
• v > 2 m/s (6.56 ft/s): for fluids producing buildup (e.g. wastewater sludges)
For an overview of the measuring range full scale values, see the "Measuring range" section
Pressure loss
• No pressure loss occurs if the sensor is installed in a pipe with the same nominal diameter.
• Pressure losses for configurations incorporating adapters according to DIN EN 545 (→ 27)
System pressure
• Never install the sensor on the pump suction side in order to avoid the risk of low pressure, and
thus damage to the liner.
• Furthermore, install pulse dampers if reciprocating, diaphragm or peristaltic pumps are used.
• For information on the liner's resistance to partial vacuum (→ 33)
• For information on the measuring system's resistance to vibration and shock (→ 30),
A0015594
Vibrations
In the event of very strong vibrations, the pipe and sensor must be supported and fixed.
It is also advisable to mount the sensor and transmitter separately.
For information on the permitted resistance to vibration and shock (→ 30), (→ 30)
L
A0016266
26
Measures to prevent vibration of the device
L
> 10 m (33 ft)