Proline Promag L 400
Hauser
15
Signal transmission 0-20 mA/4-20 mA HART with additional outputs and inputs
Order code for
"Output" and
"Input"
Terminal numbers
Output 1
Output 2
Output 3
Input
26 (+)
27 (-)
24 (+)
25 (-)
22 (+)
23 (-)
20 (+)
21 (-)
Option
H
•
4-20 mA HART
(active)
•
0-20 mA
(active)
Pulse/frequency
output
(passive)
Switch output
(passive)
-
Option
I
•
4-20 mA HART
(active)
•
0-20 mA
(active)
Pulse/frequency/
switch output
(passive)
Pulse/frequency/
switch output
(passive)
Status input
EtherNet/IP connection version
The sensor can be ordered with terminals or a device plug.
Connection methods available
Possible options for order code
"Electrical connection"
Outputs
Power
supply
Terminals
Terminals
• Option
A
: coupling M20x1
• Option
B
: thread M20x1
• Option
C
: thread G ½"
• Option
D
: thread NPT ½"
Device plug
Terminals
• Option
L
: plug M12x1 + thread NPT ½"
• Option
N
: plug M12x1 + coupling M20
• Option
P
: plug M12x1 + thread G ½"
• Option
U
: plug M12x1 + thread M20
1
2
L+/L
L-/N
1
2
A0020428
1
Supply voltage (wide range power unit)
2
EtherNet/IP
Supply voltage
Order code for "Power supply"
Terminal numbers
1 (L+/L)
2 (L-/N)
Option
L
(wide range power unit)
AC100 to 240 V
AC/DC24 V
Содержание Promag L 400