JB-70 Dental X-Ray System 

User Manual 

00-02-1568 Rev A 

 

 

 

 

 

 

 

 

 

Progeny Dental 

1407 Barclay Boulevard 

Buffalo Grove, Illinois  60089  U.S.A. 

Phone: (888) 924-3800 Fax: (847) 459-5175 

WWW.progenydental.com 

Содержание I7016

Страница 1: ...JB 70 Dental X Ray System User Manual 00 02 1568 Rev A Progeny Dental 1407 Barclay Boulevard Buffalo Grove Illinois 60089 U S A Phone 888 924 3800 Fax 847 459 5175 WWW progenydental com...

Страница 2: ...Progeny Dental 2002 Patents Pending...

Страница 3: ...JB 70 Dental X Ray System 9 Operator Panel Controls 9 Taking an X Ray 10 Using the 12 in Cone 30 A2033 11 Recommended Maintenance 12 Regular Maintenance 12 Cleaning and Disinfecting 12 Checking Syste...

Страница 4: ...Table of Contents JB 70 ii...

Страница 5: ...e Articulating Arm provides the articulation support for the Tubehead and the reach and coverage of the Tubehead to the patient The Articulating Arm allows smooth movement for precise positioning and...

Страница 6: ...General Information JB 70 4 Figure 1 Component Diagram...

Страница 7: ...ment UL 2601 1 File Number E181750 Classified by Underwriters Laboratories Inc with respect to electrical shock fire and mechanical hazards only in accordance with UL 2601 1 and CAN CSA C22 2 NO 601 1...

Страница 8: ...ollowing regulatory documents apply 21CFR UL 2601 1 IEC 60601 1 2 IEC 60601 1 3 IEC 60601 2 7 IEC 60601 2 28 IEC 60601 2 32 Medical Device directive ISO 13485 EN46001 Declaration Progeny Inc declares...

Страница 9: ...tional Electrical code or VDE standards concerning provisions of an additional protective earth ground terminal for power supply connection Before cleaning or disinfecting this equipment must always b...

Страница 10: ...ATION SEULEMENT PAR DU PERSONNEL AUTORISE VOIR MANUEL DE L OPERATEUR WARNING X RAY THIS X RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE...

Страница 11: ...can also be adjusted using the up and down Exposure Time arrows For a table of the pre programmed exposure times set at the factory refer to the Pre programmed Exposure Times table later in this manua...

Страница 12: ...eceptor Type is selected 4 Verify that the system is set for the appropriate Patient Size The icon for the currently selected Patient Size is illuminated The change the Patient Size press the Patient...

Страница 13: ...one 30 A2033 is recommended when using the paralleling film positioning technique Using the long cone requires the use of longer exposure times which can be programmed into the system by changing an i...

Страница 14: ...roblems are resolved Regular Maintenance In the interest of equipment safety a regular maintenance program must be established This maintenance program should consist of cleaning and disinfecting as w...

Страница 15: ...is in the ON position Operator Panel Controls With the power switch located at the base of the Control Unit in the ON position verify that a time indicates on the Operator Panel Also check the functi...

Страница 16: ...programmed exposure times make sure that the proper image receptor type is selected 2 Verify proper operation of the film processor No X Ray If no x ray is produced check the following 1 Verify that t...

Страница 17: ...an be modified using the procedure that follows the table Times for 8 inch 20 cm Cone in Seconds 60 Hz 50 Hz times shown in parentheses Digital Receptor D speed Film E F Speed Film Tooth Selection Adu...

Страница 18: ...ld Incisor 0 133 0 160 0 100 0 120 0 417 0 500 0 200 0 240 0 233 0 280 0 117 0 140 Bicuspid 0 167 0 200 0 083 0 100 0 533 0 640 0 267 0 320 0 333 0 400 0 167 0 200 Bitewing 0 200 0 240 0 100 0 120 0 4...

Страница 19: ...e time is to be changed a Use the Image Receptor Type button to select the receptor type b Use the Tooth Selection button to select a tooth c Use the Patient Size button to select the patient size 3 U...

Страница 20: ...50 Hz operation Timer Accuracy 34 ms Source to Skin Distance 8 inch 20 cm 12 inch 30 cm Radiation Output Approximately 800 mR at the following sets 70 kVp 8 mA 1 0 sec 20 cm SID Minimum Inherent Filtr...

Страница 21: ...adiation produced during this time is small it is useful and contributes to forming the image on the film When the exposure time starts the first several x ray pulses will be high kVp about 80 low mA...