4. Putting the blood-pressure monitor into operation
4.1. Inserting the batteries
After you have unpacked your device, first insert the batteries. The battery compartment is located on the
underside of the device (see illustration).
a) Remove cover as illustrated
b) Insert the batteries (4 x size AA 1,5 V), thereby
observing the indicated polarity.
c) If the battery warning appears in the display, the
batteries are empty and must be replaced by new
ones.
Attention!
• After the battery warning appears, the device is
blocked until the batteries have been replaced.
• Please use «AA» Long-Life or Alkaline 1.5 V Batteries.
The use of 1.2 V Accumulators is not recommended.
• If the blood-pressure monitor is left unused for long
periods, please remove the batteries from the device.
Functional check:
Hold on the O/I button down to test all the display elements. When functioning correctly all segments must
appear.
4.2. Tube connection
•
Cuff AC-1
(1-tube cuff + T-connector):
Set up (or check) the tube connections bet-
ween cuff, instrument and pump ball as sho-
wn in the diagram.