Page 15
If the ZUA 2000 is heavily soiled, it has to be entirely disassembled and cleaned:
•
Disassemble as described above, then:
•
Remove the adjusting wheel (1) and the fixing screw (2) from the application
blade (6) completely.
•
Disassemble the application blade in its 3 parts, the upper part and the lower
part containing the steel reinforcement.
If the steel reinforcement is soiled so heavily that it does not fall out
automatically, fasten a screw halfway into the thread of the steel
reinforcement and use it to pull out the reinforcement.
If the device has to be disinfected, do not use disinfectants that contain
sodium hydroxide.
•
Clean the parts with water or with a suitable solvent.
Please take care not to damage the application surface of the application
blade and the lower edges of the side plates, otherwise inexact or non-
uniform applications may occur.