Measuring Process 

17

5.5 

Measure with Grid

Model SCANLOG only
This function displays concrete covers over a large area 
in grey scale or colour shades. The smallest concrete 
cover measured in a grid field is displayed.

Settings

See also «Settings», page 6.
•  Enter the bar diameter of the first layer.
•  Enter the object number.
•  Enter the bar spacing under the menu option 
«Corrections» 

  → «Neighb. Bar Corr.» if the correction is required.

The «2-Layer Correction» can not be used by this 
function.
•  Select the size of the grid field («MENU» 

 «Basic

  setups» 

 «xy-grid»).

The grid should usually be greater than the division of 
the first layer (e.g. +50 %) to ensure that at least one 
rebar is within a grid field.

•  Set the range in which the concrete covers are to be
  displayed in grey scale («MENU» 

 «Basic setups»

 

 «Display»).

Change the grey scale range to improve the contrast of 
the displayed object (see Fig. 5.12).

Fig. 5.12  Adjusting grey scale range

The range can be adjusted as desired even after 
measuring.
•  Select the «Measure with Grid» function.
•  Pressing the START key calls the measuring display:

Fig. 5.13  «Measure with Grid» measuring display

Содержание Profometer 5+

Страница 1: ......

Страница 2: ...Bedienungsanleitung Operating Instructions Mode d emploi Bewehrungs Suchsystem Rebar Detection System Syst me de d tection d armatures Modell S SCANLOG...

Страница 3: ...tion Drive Aliquippa PA 15001 USA Phone 1 724 512 0330 Fax 1 724 512 0331 info usa proceq com Proceq Asia Pte Ltd 12 New Industrial Road 02 02A Singapore 536202 Republic of Singapore Phone 65 6382 396...

Страница 4: ......

Страница 5: ...English...

Страница 6: ......

Страница 7: ...Settings 6 Bar Diameter 6 Object Number 7 Limit Value 7 Corrections 7 Language 8 Basic Setups 8 Data Output 8 Measure with Statistics 9 Scanning Bars 9 Measure with Grid 9 5 Measuring Process 9 Measur...

Страница 8: ...uses Failure to use the rebar locator in accordance with its designated use Incorrect startup and operation of the rebar locator Unauthorized structural modifications to the rebar locator Serious dama...

Страница 9: ...covers Storing individual cover values and statistical evaluation Determining the bar diameters 2 2 Model SCANLOG Indicating device ScanCar Universal probe Fig 2 2 Model SCANLOG The SCANLOG rebar loca...

Страница 10: ...tive measuring range is displayed in the measuring display see Fig 5 1 The large measuring range should only be used if the concrete cover is larger than is shown in the intersection point of the curv...

Страница 11: ...d 16 mm concrete cover s 55 mm and smallest bar spacing a 70 mm 2 4 ScanCar Fig 2 5 ScanCar with integrated universal probe For the SCANLOG model the ScanCar is required to perform the CyberScan and...

Страница 12: ...7 and Fig 5 13 When the measuring display is visible the backlight of the LCD can be turned on and off by pressing the key for more than 2s If no display appears replace the batteries 4 Settings The...

Страница 13: ...ar diameter in close spaced parallel bar arrangements There are structures where the bar spacing is smaller than indicated in Tab 5 2 When measuring these bars the displayed value of the concrete cove...

Страница 14: ...on page 6 Warning By changing the unit from mm to inch or vice versa existing measured and stored values are falsified In order to avoid confusing results the user must clear the memory after changin...

Страница 15: ...page 6 Enter the bar diameter If the bar diameter is unknown enter 16 mm Enter the object number Limit value To avoid confusions set the limit value to 0 during the measuring process Enter the value r...

Страница 16: ...ars always start the location with the first layer If the rebars of the first layer are too close to each other it is likely to be impossible to lo cate the bars of the second layer See also Universal...

Страница 17: ...percentage of the measured values which are below the entered limit value is displayed Pressing END again completes the measurement series and starts a new one with an object number that is increment...

Страница 18: ...d as described under Storing the measured values page 11 Measuring the diameter Welded Reinforcing Meshes page 14 5 2 Detecting Insufficient Concrete Cover This function can be used for the following...

Страница 19: ...ll be too high To measure the bar diameter in the first and second layer the minimum spacings a and b as shown in Tab 5 2 page 13 are required Select the Measure w Statistics function Carry out the RE...

Страница 20: ...e menu With these settings the concrete cover of those rebars found to be situated close to one another can be deter mined precisely Only concrete covers can be stored under the selected object number...

Страница 21: ...Neighb Bar Corr if the correction is required The 2 Layer correction cannot be used by this function Enter the size of the measuring area MENU Basic setups Scan Area Select the measuring function Scan...

Страница 22: ...suring display Press the END key if the result must be deleted 16 Measuring Process 2007 Proceq SA Measuring procedure Start with the END key The following display appears Fig 5 8 Scanning display Det...

Страница 23: ...rection can not be used by this function Select the size of the grid field MENU Basic setups xy grid The grid should usually be greater than the division of the first layer e g 50 to ensure that at le...

Страница 24: ...i cated in meters or feet One page contains 16 x 15 measured values that are displayed in grey scale and stored The memory has a sto rage capacity of 500 pages The number of the remaining free pages i...

Страница 25: ...edge of the measuring area y axis is completed the cursor jumps to below the last field and the arrow direction changes automa tically The indicating device operates in the same way if the travel path...

Страница 26: ...Select All object numbers used are listed 5 6 2 Object Display An object which has been selected from the list is called from the memory and displayed 5 6 3 Object Print The object can be printed wit...

Страница 27: ...ta carrier For data transfer to the RS232 interface use the transfer cable Art No 330 00 456 If only a USB interface is avail able an additional RS232 USB adapter Art No 390 00 540 must be used Inform...

Страница 28: ...a clean dry cloth after use Clean dirty input sockets and connectors with a clean dry brush 6 2 Performance Check Check the cables for damage All measuring functions can be checked with the test block...

Страница 29: ...cl protec tive film Probe cable 1 5 m ScanCar probe carriage with Option path measuring cable 1 55 m Interface converter serial Option parallel incl 2 0 m cable Transfer cable 1 5 m USB memory stick f...

Страница 30: ...lues can be divided into 60 objects Graphics LCD 128x128 pixels with backlight Interface RS 232 Software for direct printing ProVista software for data transfer to PC and editing Power supply with 6 b...

Отзывы: