Cleaning and disinfection
Cleaning the unit
Clean the dust from the main unit by wiping it with a disposable cloth after being moistened
with water or light-detergent solution.
Other parts of the unit may be cleaned with a light-detergent solution in warm or hot
water.
Unit disinfection procedure
All parts of the nebuliser can be disinfected with 70% ethylalcohol (medical grade alcohol).
Some parts of the unit may be damaged if you use disinfectants based on:
phenols,
alkyl amine compounds,
halogenreleasing compounds,
strong organic acids,
oxygen-releasing compounds.
Wipe disinfection
The surface of the unit and its parts are disinfected:
by wiping the outer surface of the nebuliser unit and other parts with cotton wool
slightly moistened with alcohol.
do not allow contact between the disinfectant and the inside of the unit .
General advice
• Do not wipe the main unit with volatile chemicals, such as benzene or thinners.
• Do not damage the surface of the vibrator.