4.5   Turning the Device On

The Micro-Z Mini™ is powered on by pressing
and releasing the On [|] / Active button for about
2 seconds until the display lights up with P1. Only
the “P” will be fl ashing.  The Unit should be
connected to the garment electrodes and the
garment should be worn at this point.

4.6   Protocol Selection

When the unit is powered on, select desired
protocol P1 or P2 based on your clinician’s
instructions (refer to Step 5 for more details). 
Protocol selection is made by repeatedly pressing
the On [|] / Active button until the desired Protocol
is shown on the far left of the display, only the “P”
will be fl ashing.

4.7   Treatment Intensity

Two buttons on the device control intensity

Increase

 (

) and 

Decrease 

(

)

The intensity is always at fi ve when the device is
powered on or when the batteries have been
removed and replaced.
 Press 

the 

Increase

 button to set the

protocol and intensity level.  The “P” will stop
fl ashing and as you continue to press the button,
the intensity will increase and the LCD will
display the numeric value. The intensity will

increase in increments of 5, up to the maximum
intensity of 100.  To decrease the intensity, press
the 

Decrease 

(

) button.

 Note:

 The 

Increase

 and 

Decrease

 buttons

lock out after 20 seconds of nonuse so that
stimulation cannot be inadvertently changed
during treatment.
 
 

To reactivate the intensity controls, press

and release the 

On 

[|]

 / Active

 button.  Intensity

controls will remain active for 20 seconds after the
last button was pressed.

4.8   Powering the Device Off

The Micro-Z Mini™ will automatically power off
when treatment is completed.

The Micro-Z Mini™ can be powered off manually
at any time.  To manually power off, press the On
Active button for about 4 seconds and the device
will turn off.  Or you may press and hold the
decrease button until the device reads 0, release
it and press it again and the device will turn off.

Содержание Micro-Z MINI

Страница 1: ...ease read all instructions carefully before use CAUTION Federal law USA restricts this device to sale by or on the order of a physician Rev 8 2007 P N 0 61 MZMMAN 0 1 Manufactured for P O Box 40 Oakwo...

Страница 2: ...Device 4 2 Electrode Placement 4 3 Powering the Device 4 4 Connecting the Lead Wires 4 5 Turning the Device On 4 6 Protocol Selection 4 7 Treatment Intensity 4 8 Powering the Device Off 5 0 SETTING CL...

Страница 3: ...a full variety of frequencies time settings and delivery schedules Designed for use with our patented garment electrodes wearable electrodes that cover a large surface area providing total stimulatio...

Страница 4: ...ices during pregnancy TENS is not effective for pain of the central origin This includes headache TENS devices should be used only under the continued supervision of a physician or qualified professio...

Страница 5: ...veness is highly dependent upon patient selection by a person qualified in the management of pain PRECAUTIONS Continued 2 4 2 NMES 1 Precautions should be taken in the presence of A tendency to hemorr...

Страница 6: ...S 1 Micro Z Mini Device A Intensity Control Increase 2 Two 2 lead wires B Intensity Control Decrease One 1 red C On Active Switch One 1 black D Micro Z Mini 3 One 1 AAA Battery E LCD Display 4 Velcro...

Страница 7: ...ith the above cautions in mind Follow all instructions for the electrode use 4 3 Powering the Device The Micro Z Mini is powered by one 1 AAA battery Remove the battery compartment cover from the back...

Страница 8: ...the Increase button to set the protocol and intensity level The P will stop flashing and as you continue to press the button the intensity will increase and the LCD will display the numeric value The...

Страница 9: ...l run the display will show time remaining in hours and minutes or minutes and seconds During the Dwell time of protocol 2 P2 the intensity setting will flash 5 2 Clinician Value Details The Micro Z M...

Страница 10: ...not enough voltage to sustain therapy 7 0 CARE AND CLEANING Do not store the Micro Z Mini with the battery installed Battery acid causes irreparable damage which is not covered by the warranty When n...

Страница 11: ...defective parts provided the device has been properly packaged and returned postage prepaid to the manufacturer This warranty is rendered void if damage to the device results from mishandling misuse a...

Отзывы: