4.0   INSTRUCTIONS FOR USE

   4.1  Unpacking the Device

Remove the Micro-Z Mini™ System
Components from the packaging. Verify that all
components listed in Section 3.0, System 
Components, are present. The garment
electrodes and Conduct Mist™ spray may be 
packaged separately.

4.2   Electrode Placement
        WARNING!

  

 

Do not place an electrical stimulator across a

   patient's heart or transthoracically.

 

Do not stimulate over the carotid sinus nerves.

 

Severe spasm of the laryngeal or pharyngeal

   muscles may occur when electrodes are placed
   over the neck or mouth area. Contractions may
   be strong enough to close the airway or cause
   diffi culty in breathing.

 Do not apply electrical stimulation

   transcerebrally.

 Do not use electrical stimulation over

   swollen, infected or infl amed areas, or
   skin eruptions such as phlebitis,
   thrombophlebitis, or varicose veins. Place 
   electrodes according to prescribed treatment
   with the above cautions in mind. Follow all
   instructions for the electrode use.

4.3   Powering the Device

The Micro-Z Mini™ is powered by one (1) AAA
battery. Remove the battery compartment cover
from the back of the device.  Install the battery
according to the illustration inside the battery
compartment making certain the positive terminal
of the battery aligns with the ‘+’. Replace the
battery compartment cover.  

4.4   Connecting the Lead wires

The lead wires determine the polarity of the current 
and are coded accordingly; black is negative and 
red is positive. Connect the lead wire snaps to the
Micro-z Mini™ by inserting the snap stud end into 
the inside position of the slot and sliding toward 
the outside position to lock in place. The red lead 
wire is placed on the right hand side and black on 
the left. Connect the garment electrodes to the 
lead wires by inserting the pin ends into the 
garment electrode.

Garment Electrode

Please read the product insert accompanying the
garment electrode.  Connect the Silver-Thera™ or
Electro-Mesh™ garment electrode to the lead
wires using the female snaps.

 

Содержание Micro-Z MINI

Страница 1: ...ease read all instructions carefully before use CAUTION Federal law USA restricts this device to sale by or on the order of a physician Rev 8 2007 P N 0 61 MZMMAN 0 1 Manufactured for P O Box 40 Oakwo...

Страница 2: ...Device 4 2 Electrode Placement 4 3 Powering the Device 4 4 Connecting the Lead Wires 4 5 Turning the Device On 4 6 Protocol Selection 4 7 Treatment Intensity 4 8 Powering the Device Off 5 0 SETTING CL...

Страница 3: ...a full variety of frequencies time settings and delivery schedules Designed for use with our patented garment electrodes wearable electrodes that cover a large surface area providing total stimulatio...

Страница 4: ...ices during pregnancy TENS is not effective for pain of the central origin This includes headache TENS devices should be used only under the continued supervision of a physician or qualified professio...

Страница 5: ...veness is highly dependent upon patient selection by a person qualified in the management of pain PRECAUTIONS Continued 2 4 2 NMES 1 Precautions should be taken in the presence of A tendency to hemorr...

Страница 6: ...S 1 Micro Z Mini Device A Intensity Control Increase 2 Two 2 lead wires B Intensity Control Decrease One 1 red C On Active Switch One 1 black D Micro Z Mini 3 One 1 AAA Battery E LCD Display 4 Velcro...

Страница 7: ...ith the above cautions in mind Follow all instructions for the electrode use 4 3 Powering the Device The Micro Z Mini is powered by one 1 AAA battery Remove the battery compartment cover from the back...

Страница 8: ...the Increase button to set the protocol and intensity level The P will stop flashing and as you continue to press the button the intensity will increase and the LCD will display the numeric value The...

Страница 9: ...l run the display will show time remaining in hours and minutes or minutes and seconds During the Dwell time of protocol 2 P2 the intensity setting will flash 5 2 Clinician Value Details The Micro Z M...

Страница 10: ...not enough voltage to sustain therapy 7 0 CARE AND CLEANING Do not store the Micro Z Mini with the battery installed Battery acid causes irreparable damage which is not covered by the warranty When n...

Страница 11: ...defective parts provided the device has been properly packaged and returned postage prepaid to the manufacturer This warranty is rendered void if damage to the device results from mishandling misuse a...

Отзывы: