9
10 pps for 15 minutes for a total treatment of 30
minutes. P2 is programmed at 80 pps for 10
minutes and 8 pps for ten minutes totaling 20
minutes of treatment for each hour of the 8-hour
routine.
6.0 Battery Care – Low Battery
1.1 Battery Care
The Micro-Z Mini™ uses one AAA battery. To
replace the battery, open the battery compartment
at the back of the device and insert a new battery.
Be sure the ‘+’ symbol on the battery lines up with
the ‘+’ symbol in the compartment.
6.2 Low Battery
When the battery is low, a battery symbol will fl ash
on the far right of the display. Replace the battery
at the fi rst indication of low battery life. The device
will turn off automatically if there is not enough
voltage to sustain therapy.
7.0 CARE AND CLEANING
• Do not store the Micro-Z Mini™ with the battery
installed. Battery acid causes irreparable
damage, which is not covered by the warranty.
• When not in use, make certain the device is
turned off.
• Turn the power off when cleaning the device.
Do not immerse the device in liquid. Avoid
spilling liquids on the Micro-Z Mini™.
• The surface of the Micro-Z Mini™ may be wiped
with a soft cloth or sponge dampened with a mild
soap solution. Avoid caustic cleansers.
8.0 TROUBLE-SHOOTING
8.1 Device will not turn on
• Check for proper battery installation
• Replace battery.
8.2 No Stimulation:
• Check lead wires for proper connection onto the
device.
• Check for proper patient electrode application
• Apply conductive spray or cream to the area to
be stimulated
• Check treatment time for expiration.
Dwell
Time
Total Time
Segment
1 Time
Segment
2 Time