Fault finding
Should problems arise with the use of the hoist, review the following
chart. Find the fault and complete the recommended solution. If the
fault is not found and/or the solution does not correct the problem,
contact your local Prism Medical UK authorised Service Provider for
service immediately.
Problem
Solution
Hoist cannot raise/lower,
control box is not clicking.
•
Hand control is not properly plugged in.
•
Hand control is defective.
•
Control box is defective.
Hoist cannot raise/lower,
control box is clicking.
•
Lifting actuator cord (1) is not fully
plugged into slot (1) on the control box.
•
Lifting actuator is defective.
Emergency Lower button on
control box does not lower.
•
Lifting actuator cord (1) is not fully
plugged into slot (1) on the control box.
•
Lifting actuator is defective.
•
Control box is defective.
Hoist cannot spread or close
legs, control box is not
clicking.
•
Hand control is not properly plugged in.
•
Hand control is defective.
•
Control box is defective.
Hoist cannot spread or close
legs, control box is clicking.
•
Leg spreading actuator cord (2) is not fully
plugged into slot (2) on the control box.
•
Leg spreading actuator is defective.
Cannot move hoist.
•
Rear castor brake is engaged.
•
Castors are defective.
Cannot tighten Mast into
Base tube.
•
Mast locking assembly are defective,
replace with new mast locking assembly .
Содержание A205
Страница 1: ...Aluminium Mobile Hoist A205 User Manual ...
Страница 39: ...A205 Mobile Hoist User Manual Page 39 of 44 Part No 999003 Rev 2 Nov 2017 ...
Страница 40: ...A205 Mobile Hoist User Manual Page 40 of 44 Part No 999003 Rev 2 Nov 2017 ...
Страница 41: ...A205 Mobile Hoist User Manual Page 41 of 44 Part No 999003 Rev 2 Nov 2017 ...
Страница 42: ...A205 Mobile Hoist User Manual Page 42 of 44 Part No 999003 Rev 2 Nov 2017 ...
Страница 43: ...A205 Mobile Hoist User Manual Page 43 of 44 Part No 999003 Rev 2 Nov 2017 User Notes ...