Connect  the  NPC-D-6110  controller  to  the  computer  using  an  RS232C  cable  from  the  serial  port 
on  the  computer  to  the  SYNC  IN/OUT  port  on  the  NPC-D-6110.

 

A  gender  changer  adapt

e

r  is 

required to connect the RS232C cable to the sync port. The adapt

e

r required is a 9-pin male to male 

adapter as shown.   

On ‘power up’, the controller will always move the stage across its range to

 

carry out auto-calibration. 

It is important to ensure that there is sufficient clearance between the lens, sample and 
illumination to allow this to take place.  

If the temperature or load changes significantly the stage 

can 'clip' at one end of the travel.  Should this happen restart your system to allow auto-calibration. 

Once installed navigate to Device Manager and select the USB to Serial Port COM setting. Make a note of the COM port used. 

Open NS-Elements and navigate to the manage device section; install devices as shown below and select 
Prior ProScan III/

OptiScan III

.

 

CONNECTING THE CONTROLLER TO THE COMPUTER

 

DEVICE MANAGER

9 pin male to male 

adaptor

 