Connect the NPC-D-6110 controller to the computer using an RS232C cable from the serial port
on the computer to the SYNC IN/OUT port on the NPC-D-6110.
A gender changer adapt
e
r is
required to connect the RS232C cable to the sync port. The adapt
e
r required is a 9-pin male to male
adapter as shown.
On ‘power up’, the controller will always move the stage across its range to
carry out auto-calibration.
It is important to ensure that there is sufficient clearance between the lens, sample and
illumination to allow this to take place.
If the temperature or load changes significantly the stage
can 'clip' at one end of the travel. Should this happen restart your system to allow auto-calibration.
Once installed navigate to Device Manager and select the USB to Serial Port COM setting. Make a note of the COM port used.
Open NS-Elements and navigate to the manage device section; install devices as shown below and select
Prior ProScan III/
OptiScan III
.
CONNECTING THE CONTROLLER TO THE COMPUTER
DEVICE MANAGER
9 pin male to male
adaptor