background image

PriorLux

 

 

7

 

G. Oil 

Immersion 

Objectives Only 

For oil immersion lenses, make sure that a part of the specimen is in the centre of the 

field of view as seen with the last dry objective used. Rack the nosepiece up slightly, 

swing the dry lens to one side, rotate the 100x oil immersion lens over the specimen 

ensuring that it clears the cover-slip. If there is insufficient clearance, rack the 

nosepiece up further. Lower the lens close to the slide and then rotate it slightly to 

one side. Place a drop of immersion oil (free of air bubbles) on the specimen and 

gently rotate the lens into the drop. Using the fine focus knob, focus the image. 

Proceed as before through steps B to F. Finally, set the focus stop to ensure that the 

front lens of the objective does not touch the slide. 

Remove residual oil from the objective after use. 

 

 

 

 

8. Cleaning 

Objectives 

It is critical that the front lens of each objective is kept clean and free of contamination.  

Any dust or dried immersion oil will seriously affect the image quality attainable with that 

objective.  If contamination is suspected then the easiest way to confirm this is by removing 

the objective lens and examining the front lens using the eyepiece.  To do this take out one 

of the eyepieces, turn it around so that you are looking the wrong way through it and move 

it towards the front of the objective until you can focus on the front lens.  This will clearly 

show any contamination.  To remove dirt and oil a lens cloth, lens tissue or cotton bud 

dampened with industrial alcohol can be used.  A spiral motion starting from the centre of 

the lens moving to the outside is the best way of achieving a thoroughly clean surface

Содержание PriorLab

Страница 1: ...PriorLux 1...

Страница 2: ...king The PriorLab PriorLux is shipped in protective bags within a pre formed container Each component should be carefully unpacked and checked cutting rather than tearing the plastic bags The head bin...

Страница 3: ...sion control and focus stop Co axial fine and coarse adjustment with tension control and focus stop Viewing Head Interpupillary distance 55 75mm Interpupillary distance 55 75mm Kohler Illumination 12V...

Страница 4: ...4 Prior Scientific 4 Component Parts...

Страница 5: ...ng on the power reduce the lamp intensity control to its minimum setting After switching on the intensity should be increased to a comfortable level This procedure should be reversed before switching...

Страница 6: ...ct centration Figure 1 Pictures C D E Fully open the field diaphragm F Fully open the condenser diaphragm and then slowly close it until you see the contrast within the image increase If you then remo...

Страница 7: ...the objective does not touch the slide Remove residual oil from the objective after use 8 Cleaning Objectives It is critical that the front lens of each objective is kept clean and free of contaminat...

Страница 8: ...d select the correct condenser position 0 for brightfield 10 for the 10x objective 20 for the 20x objective etc Follow the steps in section 7 to ensure that K hler illumination is set up properly Remo...

Страница 9: ...trast image is poor then check that the condenser is properly focused and centred and then ensure that the rings are correctly aligned 10 Dark Ground The microscope may be used for dark ground observa...

Страница 10: ...the c mount adapter and insert the adapter with the attached camera into the top of the photo tube on the trinocular head C Tighten the screw to secure the assembly D Connect camera to a PC framegrab...

Страница 11: ...PriorLux 11...

Страница 12: ...its back to gain access to the base plate D Loosen the screw which holds the lamp cover to the base plate E Open the lamp cover to expose the bulb holder F Remove the old bulb and replace it with the...

Страница 13: ...following symbols have been used on this microscope These symbols are found next to the bulb access door on the underside of the instrument Warning High voltage disconnect power supply before changing...

Страница 14: ...use part 1 General requirements EN61326 1997 A1 A2 A3 Electrical equipment for measurement control and laboratory use EMC requirements Class B emmisions EN61326 1997 A1 A2 A3 Electrical equipment for...

Страница 15: ...PriorLux 15 17 Specification subject to change without notification Part No W2968 Issue B 10 05...

Отзывы: