7. Setting 

K

ö

hler Illumination 

A.

 

Follow the procedure in section 6 to set up the instrument for your eyesight. 

B.

 

With a specimen in sharp focus using the 10x objective, close the field diaphragm 

until it impinges on the field of view, Figure 1, Picture A. 

                   

Figure 1 

C.

 

Focus the condenser using the side mounted rack and pinion controls until the 

leaves of the diaphragm are in sharp focus, Figure 1, Picture B. 

D.

 

Using the condenser centration controls move the diaphragm into the centre of the 

field of view, it may help to open it until it is nearly touching the outside of the 

field to attain perfect centration, Figure 1, Pictures C & D. 

E.

 

Fully open the field diaphragm. 

F.

 

Fully open the condenser diaphragm and then slowly close it until you see the 

contrast within the image increase.  If you then remove an eyepiece and look 

directly down the tube from a distance of 20-30cm you should see an image similar 

to picture D in Figure 1.  The aim here is to have the “bright” area occupying 

approximately 70% of the total field.  The amount it occupies will change 

dependant on the objective lens in use.  If you have set up the instrument using the 

10x objective (recommended) then as you increase the magnification this 

diaphragm will need opening to optimise the contrast and resolution.  Often it is set 

for the objective which is either most frequently used or most critical in terms of 

resolution and left in that position

Содержание PriorLab

Страница 1: ...PriorLux 1...

Страница 2: ...king The PriorLab PriorLux is shipped in protective bags within a pre formed container Each component should be carefully unpacked and checked cutting rather than tearing the plastic bags The head bin...

Страница 3: ...sion control and focus stop Co axial fine and coarse adjustment with tension control and focus stop Viewing Head Interpupillary distance 55 75mm Interpupillary distance 55 75mm Kohler Illumination 12V...

Страница 4: ...4 Prior Scientific 4 Component Parts...

Страница 5: ...ng on the power reduce the lamp intensity control to its minimum setting After switching on the intensity should be increased to a comfortable level This procedure should be reversed before switching...

Страница 6: ...ct centration Figure 1 Pictures C D E Fully open the field diaphragm F Fully open the condenser diaphragm and then slowly close it until you see the contrast within the image increase If you then remo...

Страница 7: ...the objective does not touch the slide Remove residual oil from the objective after use 8 Cleaning Objectives It is critical that the front lens of each objective is kept clean and free of contaminat...

Страница 8: ...d select the correct condenser position 0 for brightfield 10 for the 10x objective 20 for the 20x objective etc Follow the steps in section 7 to ensure that K hler illumination is set up properly Remo...

Страница 9: ...trast image is poor then check that the condenser is properly focused and centred and then ensure that the rings are correctly aligned 10 Dark Ground The microscope may be used for dark ground observa...

Страница 10: ...the c mount adapter and insert the adapter with the attached camera into the top of the photo tube on the trinocular head C Tighten the screw to secure the assembly D Connect camera to a PC framegrab...

Страница 11: ...PriorLux 11...

Страница 12: ...its back to gain access to the base plate D Loosen the screw which holds the lamp cover to the base plate E Open the lamp cover to expose the bulb holder F Remove the old bulb and replace it with the...

Страница 13: ...following symbols have been used on this microscope These symbols are found next to the bulb access door on the underside of the instrument Warning High voltage disconnect power supply before changing...

Страница 14: ...use part 1 General requirements EN61326 1997 A1 A2 A3 Electrical equipment for measurement control and laboratory use EMC requirements Class B emmisions EN61326 1997 A1 A2 A3 Electrical equipment for...

Страница 15: ...PriorLux 15 17 Specification subject to change without notification Part No W2968 Issue B 10 05...

Отзывы: