STERILIZER Company, LLC • 6565 South 118th Street • Omaha, NE 68137• Phone (402) 344-4200 • Fax (402) 344-4242
Page 3 of 6
®
Process
Step 1
Ensure the control box is turned off by turning the yellow power
switch to the “Off” position.
Step 2
Carefully remove the top ribbon connector from the top of the
SBC196 board, noting that the red striped pin will be at the top of
the socket during reassembly.
Step 3
Carefully loosen and remove the four small nuts and nylon
washers from the four corners of the SBC196 board and place
them to one side. These will be needed later to re-assemble the
board. Note the orientation of the board. A good idea would be to
document this on a cell phone by taking a reference photograph of
the board before going any further. Because the four stand offs
have loose parts, be careful when handling the board to not loose
any parts. The nylon parts could be removed, but this is not
required.
Step 4
Grasp the edges of the SBC196 board and lift it straight up to
disengage it from the bottom board. It may be necessary to
wiggle it gently from side to side while pulling straight up to loosen
it. Make sure to carefully brace the bottom board when pulling to
ensure it does not flex too much. Be careful to avoid bending the
pins on the underside of the SBC196 board when removing it.
Step 5
Place the SBC196 board off to one side, again being careful not to
damage the pins on the bottom of the board.
Step 6
The I/O jumpers on the interface control board should now be
visible. Refer to the photograph below. Set the configuration to
match the sterilizer.
Take another reference photograph before proceeding.
NO: 258
APRIL 30, 2014
PSS500 CONTROL BOARD JUMPERS - FIELD CONFIGURATION
(CONTINUED)