PSS500 CONTROL BOARD JUMPERS - FIELD CONFIGURATION

®

Overview

The PSS500 control system uses jumpers to configure the control board for the
specific configuration of options and relays being used on a sterilizer.

The configuration is based on the sterilizer purchase order options and the
proposed purpose (Healthcare, BioPharma, Laboratory, Vivarium, etc.) for the
sterilizer.

The jumper configuration is set by PRIMUS using the electrical drawings and
specified Opto-22 relay board configuration for the sterilizer being built.

The PSS500 control board and jumper configuration are further checked during the
Factory Acceptance Testing (FAT).

Caution

Under normal conditions, there IS NO reason to adjust or re-configure
jumpers on a PSS500 control board.   Unless specifically directed to
alter the jumpers by PRIMUS, field maintenance IS NOT required on
these settings.

Returning a Control Board to PRIMUS

If a control board is returned to PRIMUS for testing due to some kind of problem, it
is important the jumpers not be altered prior to sending the board back to PRIMUS.
This will allow PRIMUS to test the board as it was used in the field.

When replacing a PSS500 control board (PRIMUS PN 200128), PRIMUS will
configure the jumpers before sending the board to the field. The respective
configuration will be based on the sterilizer’s Job Number.

Содержание PSS500

Страница 1: ... 22 relay board configuration for the sterilizer being built The PSS500 control board and jumper configuration are further checked during the Factory Acceptance Testing FAT Caution Under normal conditions there IS NO reason to adjust or re configure jumpers on a PSS500 control board Unless specifically directed to alter the jumpers by PRIMUS field maintenance IS NOT required on these settings Retu...

Страница 2: ...ot recommend configuring jumpers in the field as it can be a tedious exercise that requires strong attention to detail and knowledge of electrostatic discharge ESD practices Note ESD is the sudden flow of electricity between two electrically charged objects caused by contact or an electrical short Normally a buildup of static electricity precedes the ESD Sometimes a spark can be seen but often the...

Страница 3: ...he board before going any further Because the four stand offs have loose parts be careful when handling the board to not loose any parts The nylon parts could be removed but this is not required Step 4 Grasp the edges of the SBC196 board and lift it straight up to disengage it from the bottom board It may be necessary to wiggle it gently from side to side while pulling straight up to loosen it Mak...

Страница 4: ...S FIELD CONFIGURATION CONTINUED There are 24 jumpers located under the microcontroller board that determine whether an I O is an Input or Output These are labeled A X and correspond to Opto22 modules 0 23 as shown in the table below A B C D E F G H I J K L M N O P Q R S T U V W X 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 A shorting block on the jumper sets the I O as an Output ...

Страница 5: ...ctrical diagram provided in the Operator Manual and in the electrical box shipped with the sterilizer Without this diagram it will be extremely difficult if not impossible to correctly configure the board If needed the electrical diagram can be requested from PRIMUS Be cautious when setting and removing the jumpers to avoid bending the jumper pins Once the jumpers are set it is a good idea to docu...

Страница 6: ... 11 Attach the four nylon washers and small nuts to the stand offs Step 12 Attach the ribbon cable to the SBC196 board with the red stripe at the top of the outer socket where the yellow capacitor is located Be careful not to bend any pins when re attaching the cable end Step 13 Verify all connections are secure Step 14 Return power to the control board Additional Information For further informati...

Отзывы: