Page 14
STR
: the calibration strategy of the programmed reference values. The available
strategies are:
(UP),
(DOWN),
(UP - DOWN),
(DOWN - UP),
(UP - DOWN - UP) and
(DOWN - UP - DOWN).
RP
: the strategy number of repetitions.
RGI
: the thermo-element indication range.
RGO
: the operating temperature range that corresponds to the indication range
above.
The automatic calibration begins when the option
EXEC
from menu
COM
is
selected. All the operations are automatically performed by the T-25N calibrator. The
keypad does not work until the end of the calibration.
At the end of the stabilization time, the calibrator stores the thermo-element
reading in memory and goes to the next point, in case some input signal has been
previously configured in the menu
IN
and connected to the calibrator terminals.
The
CALIBRATION
END
message appears on the display at the end of the
automatic calibration. Press
ENTER
to confirm. The results can be verified in option
VERIF
from menu
COM
.
The first message in option
VERIF
informs the calibration result, with the number
of points that have succeeded. Following, press
ENTER
to verify each one of the
readings performed by the calibrator. The keys
and
interchange 2 screens: one
that displays the block and thermo-element temperature value and the other displays a
message indicating the calibration point number and its state (
OK
or
FAIL
), besides the
error value in %.
The option
TAG
from menu
COM
can present a list of up to 4 tags that have been
downloaded or uploaded with the ISOPLAN-5 software. The key
ENTER
selects the tag
to be calibrated from the tag list. The automatic calibration starts by the option
EXEC
from menu
COM
and the calibration data can be verified in option
VERIF
. To calibrate
automatically not using ISOPLAN-5, there is a reserved place for one manual tag. The
selection of the manual tag is performed by confirming the option
TAGMAN
with the key
ENTER
.
The option
ADDRESS
selects the communication address of the T-25N calibrator.
The communication protocol used is ModBus - RTU, with no parity and baud rate of
9600. The calibrator communication with the computer can use RS-232 or RS-485, for
network option, according to the communication interface used. In order to communicate
with the ISOPLAN-5 software, configure
ADDRESS
to 1.