2.4 - Programmable Operating Mode 

Pre-configured programs can be loaded from the calibrator memory, enabling the 

programmable  operating  mode.  The  temperature  programmed  values  of  the  block  are 

used directly, with no need to enter the setpoint. 

The  display  shows 

STEPn

  beside  the  block  temperature  setpoint  value  in  the 

programmable mode. The number of the program is indicated by 

n

. Using the keys 

 

and 

,  the  programmed  values  of  temperature  setpoint  are  changed.  The  numeric 

keypad  continues  available  for  manual  selection  of  the  block  temperature  in  the  same 

way of the manual operating mode. 

The  automatic  scan  over  the  programmed  temperatures  is  implemented  by 

defining the stabilization time of the thermo-element in the block. 

The  key 

  enables  the  automatic  scan  over  the  points.  When  pressed,  the 

message 

STEPn

  gives  place  to 

0s

  and  the  calibrator  waits  for  the  stabilization  time 

configured from 1 to 9 minutes, by the keys

 1

 to 

9

. The automatic scan is disabled by 

pressing key 

 again. 

A countdown of the stabilization time is only started when the block temperature 

reaches  the  programmed  temperature  and  stabilizes  within  a  range  of  approximately 

0.05 ºC. At this time the buzzer beeps, in case it is configured to 

YES

2.5 - Automatic Operating Mode 

The  thermo-element  calibration  is  performed  in  an  automatic  way  by  the  T-25N 

calibrator. The configuration, as well as the calibration  verification is carried out by the 

calibrator itself. Also it is possible to use the  ISOPLAN-5 software and its work orders, 

like a CAC - Computer Aided Calibration. 

The independent automatic calibration, without the use of ISOPLAN-5, is planned 

in the option 

TAGMAN

 from menu

 COM

 

TAG 

EXEC 

VERIF 

  ADDRESS 

 

TAGMAN 

Before  you  start  programming,  configure  first  the  signal  that  will be read by the 

calibrator  in  the  menu

  IN

.  To  calibrate  glass  thermometers,  for  example,  there  is  no 

electric signal to be read. In this case, option 

IN

 from the menu must be configured to 

NO

 and the calibrator will ask to enter the value indicated by the thermometer, at the end 

of the stabilization time of each calibration point. 

The data for an automatic calibration concern: 

 

TAG

: the thermo-element tag identification. 

 

SP

: the block temperature reference values for the calibration (calibration points). 

 

TOL

: the maximum tolerance for the thermo-element operation. 

 

STB

:  the  stabilization  time,  in  seconds,  so  that  the  thermo-element  temperature 

indicates correctly. This timer starts just after the block reaches and stabilizes at the 

setpoint temperature. 

Содержание T-25N

Страница 1: ......

Страница 2: ...ent without removing the plug from the supply WARNING Much electromagnetic noise can cause instability to the equipment The equipment is provided with electromagnetic interference filters that protect not only the mains but also the equipment itself against noise These filters have no function if the unit is not earthed properly WARNING High temperatures are achieved in this equipment Risk of fire...

Страница 3: ...ng Mode 12 2 4 Programmable Operating Mode 13 2 5 Automatic Operating Mode 13 3 Recommendations as regards Accuracy of Measurements 15 3 1 Getting a Better Accuracy from the Dry Block 15 3 2 Recommendations to Other types of Insert 16 4 Safety Instructions 17 5 Calibrator Warning Messages 17 6 Calibration 18 6 1 Input Calibration 18 6 2 Probe Calibration 19 7 Instructions for Hardware Maintenance ...

Страница 4: ...s temperatures from 25 ºC to 125 ºC 13 ºF to 257 ºF Presents input for thermocouples thermoresistances thermoswitches Besides generating temperature it measures the signal from the sensor being calibrated Allows the control of the dry block temperature by using an external probe optional The temperature calculation uses Callendar Van Dusen coefficients Makes no use of external standard thermometer...

Страница 5: ...ric signal calibrator Independent circuitry for over temperature protection and safety Insert to choose carrying case strap and test leads included If the insert is not specified it will be provided the insert type IN01 Cup like insert with 3 4 hole hole does not reach the bottom together with a set of tiny balls made of steel used for fulfillment included 1 1 Technical Specifications Operating Ra...

Страница 6: ...2 ºF 0 1 ºC 0 1 ºF 0 1 ºC 0 1 ºF 0 1 ºC 0 1 ºF 0 6 ºC 1 2 ºF 0 4 ºC 0 8 ºF 0 2 ºC 0 4 ºF IEC 60584 IEC 60584 IEC 60584 TC E 270 to 150 ºC 454 to 238 ºF 150 to 1000 ºC 238 to 1832 ºF 0 1 ºC 0 1 ºF 0 1 ºC 0 1 ºF 0 3 ºC 0 6 ºF 0 1 ºC 0 2 ºF IEC 60584 IEC 60584 TC N 260 to 200 ºC 436 to 328 ºF 200 to 20 ºC 328 to 4 ºF 20 to 1300 ºC 4 to 2372 ºF 0 1 ºC 0 1 ºF 0 1 ºC 0 1 ºF 0 1 ºC 0 1 ºF 1 0 ºC 2 0 ºF 0...

Страница 7: ...4 06 0001 21 Probe 1 5 DIN 90º bend mini DIN connector 04 06 0007 21 Communication Interface Description Order Code RS 232 9 way D type Connector COM1 06 02 1002 00 RS 232 25 way D type Connector COM2 06 02 1004 00 RS 485 06 02 1006 00 Carrying Case Order Code 06 01 0006 00 ISOPLAN 5 Calibration Software Calibration Certificate ...

Страница 8: ...eadings are stored in the calibrator memory When powered on ON OFF key the calibrator goes through a self test routine and shows the last calibration date and the value of the battery voltage The battery voltage is constantly monitored and the low battery warning is provided After the self test is completed the display shows the starting menu IN EXEC CONF CAL COM The initial setpoint is 25 00 ºC B...

Страница 9: ...ANUAL If the internal compensation is selected the cold junction temperature value is displayed by the calibrator If the MANUAL option is selected the cold junction must be supplied by the operator After confirming the value by pressing ENTER the calibrator goes back to the operating mode RTD Selects the type of thermoresistance used Choose from PT100 NI100 CU10 and PT1000 types Choose also if the...

Страница 10: ...PRESYS Instruments T 25N Page 7 2 1 1 Input Connections Diagrams ...

Страница 11: ...sensor into the insert and plug the mini DIN connector to the external probe input EXT PROBE located on the calibrator front panel as illustrated in the figure below The temperature control with the external probe must be enabled by the PRBe option of the PRG menu More details in item 2 2 Menu CONF ...

Страница 12: ... the date and time of the calibrator Thus when the calibrator performs a calibration in the automatic mode via ISOPLAN 5 all data are registered together with date and time of occurrence Every time the calibrator is powered off the internal clock does not continue to be updated However the ISOPLAN 5 software can automatically update the calibrator date and time with the computer internal clock Oth...

Страница 13: ...INT LOW in case of 10 20 or 25 fixed steps of the range The option VARIABLE allows the user to define from 2 to 11 values of temperature setpoint not necessarily in ascending order To verify the step values of a program select the VARIABLE option and confirm the values shown in the display with the key ENTER The 10 20 and 25 options change the number of steps automatically and recalculate the valu...

Страница 14: ...rature transmitter calibration for instance because it displays the current temperature and the transmitter reading mA in the same unit Thus the error can be verified directly in ºC or ºF Select the option SC and press ENTER If no input is selected in IN the calibrator will show the SELECT INPUT FIRST message In this case go to menu IN and select the input signal type The function SC will show IN ...

Страница 15: ...CE to cancel 2 3 Manual Operating Mode The display shows the selected temperature value of the block and also the current block temperature or thermo element temperature value There are 4 ways in which the information are shown covering the calibrator input value IN the block temperature PRBi or PRBe measured with the internal or external probe respectively and the temperature setpoint SET The key...

Страница 16: ... approximately 0 05 ºC At this time the buzzer beeps in case it is configured to YES 2 5 Automatic Operating Mode The thermo element calibration is performed in an automatic way by the T 25N calibrator The configuration as well as the calibration verification is carried out by the calibrator itself Also it is possible to use the ISOPLAN 5 software and its work orders like a CAC Computer Aided Cali...

Страница 17: ...the number of points that have succeeded Following press ENTER to verify each one of the readings performed by the calibrator The keys and interchange 2 screens one that displays the block and thermo element temperature value and the other displays a message indicating the calibration point number and its state OK or FAIL besides the error value in The option TAG from menu COM can present a list o...

Страница 18: ...use of the insert below 0 C increase the setpoint to a positive temperature value withdraw the insert from the block and the sensor dry completely all these parts before continuing the calibration This procedure guarantees an accuracy next to 0 1 C In case you can accept higher values of accuracy like 0 2 C or more these cares can be left apart 3 1 Getting a Better Accuracy from the Dry Block The ...

Страница 19: ...ther than the one used to adjust the dry block for example the cup like insert fulfilled with tiny steel balls the accuracy from the manual is no longer valid since the thermal contact is different in relation to that insert However another control sensor adjustment can be performed to the new insert Do that using an external probe with high accuracy placed in the new insert and follow the steps d...

Страница 20: ...ait until they reach the ambient temperature so that the heterogeneous cooling of the parts do not jam each other 5 Calibrator Warning Messages Warning Meaning Procedure RAM ERROR READ MANUAL Problem in RAM memory Turn the calibrator off and on If the error persists send the instrument to the factory EEPROM ERROR READ MANUAL Problem in EEPROM memory Same as the previous item LOW BATTERY Level of b...

Страница 21: ...ed and once it has been filled in it will be displayed every time the calibrator is turned on Options for IN calibration are mV mA OHM CJC PROBE 6 1 Input Calibration Select the corresponding mnemonic and apply the signals presented in the tables below When calibrating inputs the display shows on the 2 nd line the value measured by the calibrator and on the 1 st line the same value is expressed as...

Страница 22: ...re sensor internal Probe RESTORE Restores the internal temperature sensor calibration parameters to the factory values To readjust the internal Probe it is necessary to compare the value indicated by the calibrator Probe and the temperature value from an external probe placed in the dry block insert The temperature of the external probe should have high accuracy and should be measured by a superth...

Страница 23: ...he T 25N temperature calibrator that can be repaired by the user Only the 6Amp fuse placed within the socket on the rear can be replaced by the user The fuse may blow due to a voltage spike in the mains or a calibrator component fault Replace the fuse once If a second fuse blows again it is because the fault is not that simple Send the calibrator to the factory for repair 8 Instructions for Fittin...

Страница 24: ...actory using an insert with an appropriate well that fits the standard sensor snugly Thus the accuracy specifications presented in this manual are not valid any more Anyway consider an increase of up to five times the accuracy values To use the tiny balls and continue achieving accurate measurements it is necessary to read from an external reference sensor tied together with the sensor to be calib...

Страница 25: ...www presyscorp com ...

Отзывы: