PRESYS
Instruments
PCON-Y17
Page 60
MEASure[:SCALar]:VOLTage[:DC]?
Description:
Voltage Input (Auxiliary) Reading.
Parameters:
---
Response:
V Reading
OUTPut:MODE <mode>
Description:
Sets the Pressure Operation Mode.
Parameters:
VENT | CONTrol | MEASure
Response:
---
3 - Maintenance
3.1. Replacing the Current Input Fuse
To replace the PCON-Y17 mA Input (Auxiliary) fuse, proceed as follows:
- Rotate the fuse holder counterclockwise (see see
Fig. 01
for DT Version or
Fig. 03
for
Rack Mounting Version
– Fuse mA input). Try to use a plastic tool to avoid damaging the
fuse holder;
- The fuse holder will be released;
- Pull the fuse holder and remove the fuse;
- Place the spare fuse (Time-Lag Type Fuse 32 mA / 250 V / 5 x 20mm).
3.2. Replacing the Power Supply Fuse
In case of blow, the 1 Amp fuse (Time Delay Type Fuse 1 A / 250 V / 5x20mm),
placed within the power supply socket on the rear can be replaced.
The fuse may blow due to a voltage spike in the mains or a calibrator component
fault. Replace the fuse once. If a second fuse blows again, it is because the fault is not
that simple. In this case, contact the
Presys
technical support.
presys