PRESYS
Instruments
PCON-Y17
Page 8
CALIBRATOR
– selects the input/output functions (Pressure Control Module, Auxiliary
Input Signals and Pressure Measurement Module), see section
2.1
.
HART
®
– optional module that allows communication with devices that have Hart
®
Communication Protocol, see section
2.2
.
TASKS
– performs calibrations automatically, see section
2.3
.
DATA LOGGER
– record measurements, enabling visualization on chart or table, see
section
2.4
.
VIDEOS
– features videos made by
Presys
to assist in the use of the calibrator, and can
also store videos made by the user, see section
2.5
.
SETTINGS
– general instrument settings, see section
2.6
.
2.1. Calibrator Menu
To select the
Pressure Control Module
and the
Auxiliary Input
functions press
the
CALIBRATOR
button from the main menu.
When first entering the
CALIBRATOR
menu, the PCON-Y17 executes a
VENT
operation and an AUTO RESET of the pressure read in the Pressure Control Module.
When it is done, it goes to
MEASURE
mode and opens the output isolation valve (see
the module layout in section
2.1.2
). The following screen is displayed.
Fig. 05
- Calibrator initialization
presys