PRESYS
|
Instruments
DMY-2011, DMY-2035 and DMY-2036
Introduction
Page 5
1.3 - Specifications
Inputs:
•
Two inputs configurable for thermocouple (J, K, T, E, R, S, under ITS-90),
Pt-100 RTD under DIN 43760, 4 to 20mA, 0 to 55mVDC, 1 to 5VDC and
0 to 10VDC. Input impedance of 250
Ω
for mA, 10M
Ω
for 5VDC and 2M
Ω
above
5VDC. Table 1 shows the temperature ranges for thermocouples and RTD and
the resolution for linear input sensors.
Input Sensor
Measuring Range Limits
Thermocouple lower
limit
°
F
higher limit
°
F
lower limit
°
C
higher limit
°
C
Type J
-184
1886.0
-120
1030.0
Type K
-346
2498
-210
1370
Type T
-418
752
-250
400
Type E
-148.0
1436.0
-100.0
780.0
Type R
-58
3200
-50
1760
Type S
-58
3200
-50
1760
RTD
2 or 3-wire Pt-100
-346.0
1256.0
-210.0
680.0*
Linear
Range
Resolution
Voltage
0 to 55mV
6
µ
V
0 to 5V
500
µ
V
0 to 10V
1mV
Current
0 to 20mA
2
µ
A
(*)including wire resistance
Table 1 - Measuring ranges for input sensors
Outputs:
•
4 to 20 mA, 1 to 5 Vdc, 0 to 10 Vdc Analog Retransmitter, with connection for
up to two optional modules galvanically isolated of 300Vac from power supply
and inputs.
•
SPDT relay for alarm rated for 3A at 220Vac, or 10A at 220Vac under order.
In this case alarm module is not connected, but soldered to the board.
Connection for up to four modules (using both analog output connectors). In
case of using one analog output, one can use three alarm modules, or
whenever one uses two analog outputs, it is possible to have up to two alarm
modules.
•
Logic signal, open collector transistor, 24 Vdc, 40 mA maximum with isolation.
•
Solid state relay rated for 2A at 250Vac with isolation.