Instrument with the display out
Check if power supply voltage is provided to terminals 23 and 24 of the
Indicator.
Verify the integrity of fuse F1 of 2.0 A placed in the Power Supply Board as
shown in figure 21. Due to its package it is necessary to check the fuse continuity in
order to detect if it is broken.
Instrument malfunction
Check if the Indicator is configured correctly by software and hardware
(internal jumpers).
Examine if the optional modules are connected in the right spots.
Check if the voltages on flat cable 1 as shown in figure 29 are close to the
values in table 12 and if they reach the CPU Board.
Test points on flat cable 1
Voltage
Between point 1(-) and point 2(+)
5V
Between point 9(-) and point 8(+)
8V
Between point 9(-) and point 1(+)
0V
Between point 9(-) and point 10(+)
- 8V
Between point 9(-) and point 13(+)
24V
Between point 12(-) and point 11(+)
5V
Table 12 - Inspection points of voltage on flat cable 1
Figure 29 - Voltage test points of the Indicator
If the cause of the problem was not discovered, the Indicator must be sent to
factory.