premiere XH-1003 Скачать руководство пользователя страница 8

 

 

6.1 Operations 

Temperature adjustment  

1.

 

Prior to use fill with water at least halfway full – use care to avoid overflowing (Note: 

type of water – tap, distilled, purified, etc. is optional based on user requirements)  

2.

 

Flip  on power switch  in back of unit, click “ON/OFF” button  and  Temperature/Set 

temperature  indicator light (see section 4.2)  will illuminate indicating the unit is 

heating/powered up. Push the “SET” button once (unit will read “Sp”). Use the “

▲” 

or “

▼” to adjust temperature to desired setting.  

 

Time adjustment  

1.

 

Once unit is set to desired temperature, push the “SET” button again (unit will read 

“St”). Use the “

▲” or “▼” to adjust time to desired setting. Push the “SET” button 

again  and allow ample time for the Lighted  Tissue Floating Bath to warm  up.  (run 

light will be solid and heat light will begin to flash  every several seconds) 

Temperature range is from room temperature to 70ºC.  Use a laboratory thermometer 

(not included) to determine actual water temperature based on dial setting and 

increase or decrease as desired.  

2.

 

Once the Tissue Floating Bath reaches your predetermined set temperature the heat 

indicator light will turn off indicating that the heater is off and unit is holding 

temperature.     

 
Making Sure Your Floating Section Is Flat 
 

1.

 

Swipe a slide underneath the floating paraffin section to pick up the section on the 

slide. After drying the bottom of the slide place your slide on the outer edge of the 

Tissue Floating Bath. This will help the paraffin section to dry, flatten out and stick 

firmly to slide. A slide warmer might be needed for large quantities of slides or 

extended drying time. 

2.

 

Between floating paraffin sections  you can use a lint-free disposable towel to wipe 

across the water to remove any leftover floating paraffin. 

3.

 

Add additional water as needed by pouring slowly into basin; allow several minutes 

for temperature to adjust and/or increase temperature control setting as needed. 

 

Содержание XH-1003

Страница 1: ...s to obtain the best results of this equipment Warning To prevent fire or shock hazard do not expose this unit to rain or moisture Unless labeled otherwise this equipment should be used with AC 110V 6...

Страница 2: ...1...

Страница 3: ...ble of Contents Section Table of Contents Page 1 Safety Page 2 Contents Specifications Page 3 Notes of Operation Page 4 Control Panel Setup Page 5 Operations Page 6 Trouble Shooting Page 7 8 Warranty...

Страница 4: ...ue Floating Bath is for histology use along with research instruction and experiments in the universities and technical secondary schools The Tissue Floating Bath is not recommended for any other uses...

Страница 5: ...dly contact your dealer promptly 3 2 Specifications Function Parameter Technical Data Environment of use Clean dry and flat surface Stable work surface Altitude 2000m Relative humidity 80 Temperature...

Страница 6: ...uctuate User may add as much water as needed to their comfort level avoid overfilling or risk of overflow Frequently stirring water can help keep temperatures consistent throughout water bath Use only...

Страница 7: ...ON OFF Power on off 5 2 Setup 1 Open box and remove Lighted Tissue Floating Bath from packaging 2 Place Tissue Floating Bath onto flat surface Look for visible damage to unit glass tray Contact deale...

Страница 8: ...rature to 70 C Use a laboratory thermometer not included to determine actual water temperature based on dial setting and increase or decrease as desired 2 Once the Tissue Floating Bath reaches your pr...

Страница 9: ...he internal temperature controller and raise the power of heater See following steps 1 Hold SET for five seconds to enter into parameter setting status 2 Press SET again until the screen shows then pr...

Страница 10: ...djust using or arrows Set the value to one half of previous value I Adjust this parameter should be the half value of previous step I s value 5 Continue pressing SET until the screen shows then adjust...

Страница 11: ...9 Notes...

Страница 12: ...ons performed by unauthorized repair technicians or damage occurring in transit If you have questions concerning this product or warranty contact the dealer from whom it was purchased For warranty ser...

Отзывы: