6
6.1 Operations
Temperature adjustment
1.
Prior to use fill with water at least halfway full – use care to avoid overflowing (Note:
type of water – tap, distilled, purified, etc. is optional based on user requirements)
2.
Flip on power switch in back of unit, click “ON/OFF” button and Temperature/Set
temperature indicator light (see section 4.2) will illuminate indicating the unit is
heating/powered up. Push the “SET” button once (unit will read “Sp”). Use the “
▲”
or “
▼” to adjust temperature to desired setting.
Time adjustment
1.
Once unit is set to desired temperature, push the “SET” button again (unit will read
“St”). Use the “
▲” or “▼” to adjust time to desired setting. Push the “SET” button
again and allow ample time for the Lighted Tissue Floating Bath to warm up. (run
light will be solid and heat light will begin to flash every several seconds)
Temperature range is from room temperature to 70ºC. Use a laboratory thermometer
(not included) to determine actual water temperature based on dial setting and
increase or decrease as desired.
2.
Once the Tissue Floating Bath reaches your predetermined set temperature the heat
indicator light will turn off indicating that the heater is off and unit is holding
temperature.
Making Sure Your Floating Section Is Flat
1.
Swipe a slide underneath the floating paraffin section to pick up the section on the
slide. After drying the bottom of the slide place your slide on the outer edge of the
Tissue Floating Bath. This will help the paraffin section to dry, flatten out and stick
firmly to slide. A slide warmer might be needed for large quantities of slides or
extended drying time.
2.
Between floating paraffin sections you can use a lint-free disposable towel to wipe
across the water to remove any leftover floating paraffin.
3.
Add additional water as needed by pouring slowly into basin; allow several minutes
for temperature to adjust and/or increase temperature control setting as needed.